Conductivity tab
Figure 5-11 Printer dialog box, conductivity tab
Disabled, timed, and percent change radio buttons
determine how often an online
conductivity reading will be printed in conductivity mode. You may select disabled, timed, or
percent change.
•
Selecting disabled disables printing of online conductivity readings.
•
Selecting timed causes a printout to occur at the interval specified in the interval text
box and disables the percent change text box.
•
Selecting percent change causes a printout to occur only when the change in
conductivity between successive analyses exceeds the value specified in the percent
change box. Selecting percent changes also disables the interval text box.
Interval text box
allows you to determine how often a printout occurs in timed mode. You
may enter a value from 1 minute to 24 hours. The default value is 1 minute.
Limit text box
allows you to enter the percentage change in online conductivity readings that
must occur between successive analyses before a printout occurs. You may select a value of
1 to 100%. The default value is 1%.
Return icon
returns you to the setup dialog box and saves the changes. Any changes are
noted in the audit trail.
5.8
Security
The security icon allows you to modify the security settings for the system. The security dialog
box allows configuration of general security features, setting and modifying user accounts,
and viewing of the audit trail.