Figure 5-9 Analog outputs setup dialog box
TOC zero scale text box
sets the TOC reading that corresponds to the lowest output for the
selected range. You may enter a value from 0 (default) to 1999 ppb. The full scale value must
be higher than the zero scale value.
TOC full scale text box
sets the TOC reading that corresponds to the highest output for the
selected range. You may enter a value from 1 to 2000 ppb (default). The full scale value must
be higher than the zero scale value.
Conductivity zero scale text box
sets the conductivity reading that corresponds to the
lowest output for the selected range. You may enter a value from 0 (default) to 19 µS/cm. The
full scale value must be higher than the zero scale value.
Conductivity full scale text box
sets the conductivity reading that corresponds to the
highest output for the selected range. You may enter a value from 1 to 20 µS/cm (default). The
full scale value must be higher than the zero scale value.
Temperature zero scale text box
sets the temperature reading that corresponds to the
lowest output for the selected range. You may enter a value from 0 (default) to 99 °C. The full
scale value must be higher than the zero scale value.
Temperature full scale text box
sets the temperature reading that corresponds to the
highest output for the selected range. You may enter a value from 1 to 100 °C (default). The
full scale value must be higher than the zero scale value.
On error, return drop down list box
determines the output that the analyzer produces when
an alarm condition occurs. You may select 2 mA output (default), last output, or 22 mA output.
Return icon
returns you to the setup dialog box and saves any changes. Any changes are
noted in the audit trail.