PAT700 Total Organic Carbon Analyzer – Diagnostics
Page 113
Anatel
Operator Manual
PAT700 lamp replacement
To replace a lamp, follow these steps:
1)
Touch the replace lamp icon.
2)
Select the lamp to be replaced.
3)
Touch “Run Wizard” to run the lamp replacement wizard. Touch “Replace Now” to
skip the replacement wizard. Touch “Print Instructions” to send the lamp
replacement instructions to a printer, if connected.
Read the instructions completely
as you proceed with the lamp replacement.
•
If you touch “Run Wizard”, the lamp replacement wizard provides step-by-step
instructions for replacing the lamp.
Read the instructions completely
as you proceed
with the lamp replacement.
•
If you touch “Replace Now”, the display skips the replacement wizard and asks you
to power down the analyzer. See
carefully follow the procedure
described in
Lamp replacement procedure, page 130
.
Figure 8-3 Lamp replacement dialog box
8.4
Tests tab
The tests dialog box allows you to test the following analyzer functions: RS–232 output, digital
I/O, 4-20 mA outputs, printer, plumbing systems, pump, and heat exchanger fan. See