Specifications
Specifications are subject to change without notice.
Performance
Specification
A1000
A1000XP
TOC measurement
TOC method
TOC = Total Carbon (TC) – Total Inorganic Carbon (TIC)
Linear range
0.05 to 1999 ppb as
carbon
XP mode: 0.02 to 1999 ppb as
carbon
1
Standard mode: 2.00 to
1999 ppb as carbon
Detection limit
0.05 ppb
0.02 ppb
Quantitation limit
0.17 ppb
0.07 ppb
Accuracy
0.5 ppb or ±5%
whichever is greater
±0.05 ppb (MDL to 0.999 ppb)
±0.1 ppb (1.000 to 1.999 ppb)
±0.5 ppb or 5% whichever is
greater (2 to 1999 ppb)
Specification
A1000
A1000XP
Precision
±0.5 ppb
±0.05 ppb (MDL to 0.999 ppb)
±0.1 ppb (1.000 to 1.999 ppb)
±0.5 ppb or 5% whichever is
greater (2 to 1999 ppb)
TOC conductivity
range
0.2 to 18.2 MΩ-cm
XP mode: 15.0 to 18.2 MΩ-cm
Standard mode: 0.2 to
18.2 M
Ω
-cm
TOC water
temperature range
0 to 100 °C (32 to
212 °F)
Note: Use a heat
exchanger for samples
more than 50 °C (122 °F).
XP mode: 18 to 32 °C (65 to
90 °F)
Standard mode: 0 to 100 °C
(32 to 212 °F)
Note: Use a heat exchanger for
samples more than 50 °C (122 °F).
Display resolution
X.YY (0.00 to
19.99 ppb)
X.Y (20.0 to 199.9 ppb)
X (200 to 1999 ppb)
X.YYY (0.000 to 1.999 ppb)
X.YY (2.00 to 19.99 ppb)
X.Y (20.0 to 199.9 ppb)
X (200 to 1999 ppb)
Calibration stability
Annual calibration or validation recommended
Resistivity measurement
Linear range
0.01 to 18 MΩ-cm
Resistivity accuracy
2
2% over full range (compensated)
Display resolution
Three significant figures
Temperature
accuracy
±0.5 °C (9 °F)
Temperature
compensation
Compensated to 25 °C (77 °F)
1
The instrument changes to XP mode automatically when the TOC conductivity and temperature range are within the XP mode ranges shown.
2
Resistivity and accuracy specifications are for ambient temperatures from 15–35 °C (59–95 °F).