![Hach 9586sc Basic User Manual Download Page 10](http://html1.mh-extra.com/html/hach/9586sc/9586sc_basic-user-manual_3141587010.webp)
5.2 Wiring overview
shows an overview of the wiring connections inside the controller with the high voltage
barrier removed. The left side of the figure shows the back side of the controller cover.
Note: Remove connector caps from the connectors before module installation.
Figure 2 Wiring connections overview
1
Service cable connection
5
AC and DC power
connector
2
9
Discrete input wiring
connector
2
2
4-20 mA output
2
6
Ground terminals
10
Digital sensor connector
2
3
Sensor module connector
7
Relay connections
2
4
Communication module
connector (e.g., Modbus,
Profibus, HART, optional
4-20 mA module, etc.)
8
Digital sensor connector
2
5.2.1 High-voltage barrier
High-voltage wiring for the controller is located behind the high-voltage barrier in the controller
enclosure. The barrier must remain in place except when installing modules or when a qualified
installation technician is wiring for power, alarms, outputs or relays. Do not remove the barrier while
power is applied to the controller.
5.2.2 Wiring for power
W A R N I N G
Potential Electrocution Hazard. Always disconnect power to the instrument when making
electrical connections.
2
Terminals can be removed for improved access.
10
English
Summary of Contents for 9586sc
Page 151: ...4 5 4 2 4 3 1 1 6 2 7 3 8 4 6 mm 4 9 6 8 mm 5 151...
Page 152: ...5 5 1 4 152...
Page 153: ...5 2 2 153...
Page 154: ...2 1 5 2 9 2 2 4 20 mA 2 6 10 2 3 7 2 4 Modbus Profibus HART 4 20 mA 8 2 5 2 1 5 2 2 24V 2 154...
Page 156: ...5 2 3 100 250 VAC 50 60 Hz 5A 250 VAC 5A 24 VDC 5A 156...
Page 158: ...158...
Page 159: ...5 2 5 NEMA IP NEMA 4X IP66 1 2 3 3 3 3 3 2 4 2 3 1 2 1 1 1 2 3 4 500 5 6 159...
Page 275: ...Suomi 275...
Page 336: ...4 1 4 1 1 4 1 2 336...
Page 337: ...EMC 4 1 3 IECS 003 A FCC 15 15 FCC 1 2 15 FCC 1 2 3 4 5 4 2 337...
Page 338: ...4 3 1 1 6 2 7 3 8 4 6 4 9 6 8 5 DIPA 338...
Page 339: ...5 5 1 4 339...
Page 340: ...5 2 2 340...
Page 341: ...2 1 5 2 9 2 2 4 20 2 6 10 2 3 7 2 4 Modbus Profibus HART 4 20 8 2 5 2 1 5 2 2 2 341...
Page 342: ...24 PE 100 240 24 100 240 24 18 12 AWG 80 C 176 F NEMA 4X IP66 3 18 24 24 15 20 1 2 342...
Page 343: ...1 1 L1 2 N PE 2 1 24 2 24 PE 343...
Page 344: ...5 2 3 100 250 50 60 5 100 250 5 24 5 5 2 4 5 5 NEMA IP NEMA 4X IP66 100 250 344...
Page 346: ...346...
Page 347: ...5 2 5 NEMA IP NEMA 4X IP66 1 2 3 3 3 3 3 2 4 2 3 1 2 1 1 1 2 3 4 500 5 6 347...
Page 349: ...3 4 6 Startup 333 1 2 3 4 5 7 7 1 4 349...
Page 350: ...4 1 5 BACK 2 SD 6 MENU 3 HOME 7 4 ENTER SD 7 2 5 350...
Page 382: ......
Page 383: ......