2—26
User-Entered Calibration,
continued
Note: For meters with factory-calibrated ranges or methods, Standard Calibration Adjust
(SCA) will be disabled when a user-entered method is programmed into the meter. To
turn SCA back on, restore the meter to factory default calibration.
Maximum/Minimum Displayed Value
In meters with absorbance (Abs) ranges, the maximum displayed value and
minimum displayed value is related to the value of the standards entered in a user
calibration.
Measurements that exceed the minimum or maximum standards entered in the
user calibration will return a flashing number indicating “underrange” or
“overrange”. See Error Codes (page
2—12
) for more information.
Example 1
For a calibration with the following standards:
S0=0.000
S1=1.000
Maximum Displayed Value
1.000
Minimum Displayed Value
0.000