![Hach 59530-16 Instruction Manual Download Page 33](http://html1.mh-extra.com/html/hach/59530-16/59530-16_instruction-manual_3226394033.webp)
1—33
Molybdenum, Molybdate, LR
(0.02 to 3.00 mg/L Mo)
Method 8169
For boiler and cooling tower waters
Ternary Complex Method
Measuring Hints
•
The results can be expressed as mg/L molybdate (MoO
4
2–
) or mg/L sodium
molybdate (NaMoO
4
) by multiplying the mg/L molybdenum (Mo
6+
) result by
1.67 or 2.15, respectively.
•
Wipe sample cells with a soft, dry cloth before placing in the instrument.
•
After adding reagent, a green color will develop if molybdenum is present in
the sample.
Note:
The Pocket Colorimeter™ II is designed to measure solutions contained in sample cells.
DO NOT
dip the meter in the sample or pour the sample directly into the cell holder.
Summary of Contents for 59530-16
Page 10: ...1 10...
Page 12: ...1 12...
Page 15: ...1 15 Instrument Cap Cord continued Figure 1 Attaching the Instrument Cap Cord...
Page 16: ...1 16...
Page 52: ...1 52...
Page 73: ...2 1 Section 2 Instrument Manual...
Page 74: ...2 2...
Page 79: ...2 7 Instrument Operation continued Figure 1 Battery Installation...
Page 80: ...2 8...
Page 100: ...2 28...
Page 106: ...2 34...