1—64
Zinc,
continued
1.
Press the
POWER
key to
turn the meter on.
The arrow should indicate
mg/L Zn.
Note:
See
page 2—4
for
information on selecting the
correct channel.
2.
Fill a 25-mL mixing
graduated cylinder to the
20-mL mark with sample.
3.
Add the contents of one
ZincoVer
®
5 Reagent Powder
Pillow to the cylinder.
Stopper the cylinder and
invert or gently shake until
all particles are dissolved.