2—18
User-Entered Calibration,
continued
3.
Insert the reagent blank or deionized water into the meter and cover with the
cap. Press the
ZERO/SCROLL
key. The meter will display “- - - -”, followed by
“0.000”. This initializes (zeroes) the meter.
4.
Press the
MENU
key and hold it down until the display shows “USER”, followed
by “CAL”. Press
READ/ENTER
to enter the calibration mode.
5.
In factory-calibrated meters, S0 will appear in the display.
Note: When recalibrating a factory-calibrated meter or range, RES (resolution) cannot
be changed.
6.
In uncalibrated meters or meters with ranges labeled Abs, “RES” will appear.
Press
ZERO/SCROLL
to review the current resolution (decimal placement). Press
ZERO/SCROLL
again to accept the current resolution. To change the resolution,
press
READ/ENTER,
then
ZERO/SCROLL
to change the resolution. Press
READ/ENTER
to accept the new resolution. “S0” will appear on the display.
7.
Press the
READ/ENTER
key again, then enter the blank value.
Note: Press the
READ/ENTER
key to move from digit to digit. Use the
ZERO/SCROLL
key to
change the number.
8.
After completing entry of the blank value, press the
READ/ENTER
key. The
display will show “A0”.
Summary of Contents for 59530-16
Page 10: ...1 10...
Page 12: ...1 12...
Page 15: ...1 15 Instrument Cap Cord continued Figure 1 Attaching the Instrument Cap Cord...
Page 16: ...1 16...
Page 52: ...1 52...
Page 73: ...2 1 Section 2 Instrument Manual...
Page 74: ...2 2...
Page 79: ...2 7 Instrument Operation continued Figure 1 Battery Installation...
Page 80: ...2 8...
Page 100: ...2 28...
Page 106: ...2 34...