Molybdenum, Molybdate, LR,
continued
1.
Press the
POWER
key to
turn the meter on.
The arrow should indicate
the low range channel (LR).
Note:
See
page 2—4
for
information on selecting the
correct range channel.
2.
Fill a 25-mL mixing
graduated cylinder to the
20-mL mark with sample.
3.
Add the contents of one
Molybdenum 1 Reagent
Powder Pillow to the
cylinder. Stopper the
cylinder and invert or gently
shake until all particles are
dissolved.