Connect to the digital inputs
The analyzer can receive a digital signal or contact closure from an external device that causes the
analyzer to skip a sample channel. For example, a flow meter can send a digital signal when the
sample flow is low and the analyzer skips the applicable sample channel. The analyzer continues to
skip the applicable sample channel until the digital signal stops. All sample channels cannot be
skipped. A minimum of one sample channel must be in use.
Note: If all of the sample channels have no sample, the user cannot put the analyzer in shutdown mode with the
digital inputs. To put the instrument into shutdown mode or back into operation remotely, use the optional Modbus
module and write to Modbus register 49937. Write 40007 (decimal) to put the analyzer in shutdown mode. Write
40008 (decimal) to put the analyzer back into operation.
Each digital input can be configured as an isolated TTL type digital input or as a relay/open-collector
type input. Refer to
. By default, the jumpers are set for isolated TTL type digital input (logic
low = 0 to 0.8 VDC and logic high = 2 to 5 VDC; maximum voltage 30 VDC). Refer to
on page 12 to connect the device.
Figure 12 Isolated TTL type digital input
1
Jumper (12x)
3
Isolated TTL type digital input
2
Digital input connectors
2
4
Relay/Open-collector type input
Install additional modules
Modules can be added for additional output, relay or communications options. Refer to the
documentation that is supplied with the module.
2
The analyzer uses the DIG6 channel for the level sensor. Do not move the connection and the
jumpers.
16
English
Summary of Contents for 5500sc
Page 2: ...English 3 Fran ais 18 Espa ol 35 52 2...
Page 54: ...54...
Page 56: ...1 2 3 4 5 pH 1 2 56...
Page 57: ...1 1 5 9 2 6 SD 10 3 7 1 11 4 8 12 2 1 57...
Page 58: ...4 2000 6562 2000 3 58...
Page 59: ...3 1 2 4 4 1 2 5 9 2 6 10 1 3 2 7 4 1 8 59...
Page 60: ...6 5 6 6 1 4 1 5 1 3 5 2 4 60...
Page 61: ...6 1 3 5 2 4 6 30 42 2 60 30 42 2 60 GFCI GFI PE 61...
Page 62: ...7 7 1 3 8x 2 3x 4 8x 8 8 62...
Page 63: ...9 52 9 1 4 7 2 5 8 3 4 20 6 9 0 1 63...
Page 64: ...1 0 2 18 AWG 3 10 52 3 4 10 3 1 PE 2 N 3 L1 4 1 PE 2 24 3 24 64...
Page 65: ...10 SC 6773200 sc sc Smart Probe Adapter 9321000 Smart Probe Adapter 11 65...
Page 66: ...11 1 0 2 18 AWG 2x 63 NO COM NC 4 20 4 20 4 20 2 Modbus 49937 66...
Page 67: ...40007 40008 12 0 0 8 2 5 30 63 12 TTL 1 12 3 TTL 2 2 4 2 DIG6 67...
Page 68: ...MSDS SDS 133 13 3 68...
Page 69: ......