Figure 9 Connections on the main circuit board
1
External controller connection
4
Digital inputs
7
Power out
2
Smart probe connection
5
Power in
8
Power out LED (on = power is
connected to the analyzer)
3
4–20 mA outputs
6
Power switch and LED (on =
analyzer is on)
9
Relays
Connect to power
W A R N I N G
Electrocution hazard. Use a crimp
‐
on ring terminal on the main protective earth connection.
W A R N I N G
Electrical shock and fire hazards. Make sure that the user-supplied power cord and non
‐
locking plug
meet the applicable country code requirements.
W A R N I N G
Electrocution hazard. Make sure that the protective earth conductor has a low impedance connection of
less than 0.1 ohm. The connected wire conductor must have the same current rating as the AC mains
line conductor.
N O T I C E
The instrument is used for a single phase connection only.
Cord installation:
The manufacturer recommends to use the optional cord and sealing gland. Refer
to the maintenance manual for the replacement parts list. For a customer-supplied cord, three
1.0 mm
2
(18 AWG) conductors are required with a waterproof outer jacket, and the cord must be
shorter than 3 meters (10 feet). Use a sealing type strain relief to keep the environmental rating of
the instrument. Refer to
on page 3. To connect power to the instrument, refer to
English
13
Summary of Contents for 5500sc
Page 2: ...English 3 Fran ais 18 Espa ol 35 52 2...
Page 54: ...54...
Page 56: ...1 2 3 4 5 pH 1 2 56...
Page 57: ...1 1 5 9 2 6 SD 10 3 7 1 11 4 8 12 2 1 57...
Page 58: ...4 2000 6562 2000 3 58...
Page 59: ...3 1 2 4 4 1 2 5 9 2 6 10 1 3 2 7 4 1 8 59...
Page 60: ...6 5 6 6 1 4 1 5 1 3 5 2 4 60...
Page 61: ...6 1 3 5 2 4 6 30 42 2 60 30 42 2 60 GFCI GFI PE 61...
Page 62: ...7 7 1 3 8x 2 3x 4 8x 8 8 62...
Page 63: ...9 52 9 1 4 7 2 5 8 3 4 20 6 9 0 1 63...
Page 64: ...1 0 2 18 AWG 3 10 52 3 4 10 3 1 PE 2 N 3 L1 4 1 PE 2 24 3 24 64...
Page 65: ...10 SC 6773200 sc sc Smart Probe Adapter 9321000 Smart Probe Adapter 11 65...
Page 66: ...11 1 0 2 18 AWG 2x 63 NO COM NC 4 20 4 20 4 20 2 Modbus 49937 66...
Page 67: ...40007 40008 12 0 0 8 2 5 30 63 12 TTL 1 12 3 TTL 2 2 4 2 DIG6 67...
Page 68: ...MSDS SDS 133 13 3 68...
Page 69: ......