![Hach 5500sc Installation Manual Download Page 11](http://html2.mh-extra.com/html/hach/5500sc/5500sc_installation-manual_3164814011.webp)
Figure 6 Drain lines
1
Sample bypass drain (single-
stream)
3
Sample holder drain
5
Case drain
2
Sample holder overflow
4
Chemical drain
6
Rear side of the analyzer
Electrical installation
D A N G E R
Electrocution hazard.
Use either high voltage (more than 30 V RMS and 42.2 V PEAK or 60 VDC) or low voltage (less than
30 V RMS and 42.2 V PEAK or 60 VDC). Do not use a combination of high and low voltage.
Always remove power to the instrument before making electrical connections.
Do not connect AC power directly to a DC powered instrument.
If this equipment is used outdoors or in potentially wet locations, a Ground Fault Circuit Interrupt
(GFCI/GFI) device must be used for connecting the equipment to its main power source.
Protective Earth Ground (PE) connection is required.
Use only fittings that have the specified environmental enclosure rating. Obey the requirements in the
Specifications section.
W A R N I N G
Electrical shock and/or fire hazards.
Install the instrument in accordance with local, regional and national regulations.
Externally connected equipment must have an applicable country safety standard assessment.
A local disconnect is needed for a conduit installation.
Make sure to identify the local disconnect clearly for the conduit installation.
For a cord-connected instrument, make sure to install the instrument so that the cord can be
disconnected easily from the supply socket.
English
11
Summary of Contents for 5500sc
Page 2: ...English 3 Fran ais 18 Espa ol 35 52 2...
Page 54: ...54...
Page 56: ...1 2 3 4 5 pH 1 2 56...
Page 57: ...1 1 5 9 2 6 SD 10 3 7 1 11 4 8 12 2 1 57...
Page 58: ...4 2000 6562 2000 3 58...
Page 59: ...3 1 2 4 4 1 2 5 9 2 6 10 1 3 2 7 4 1 8 59...
Page 60: ...6 5 6 6 1 4 1 5 1 3 5 2 4 60...
Page 61: ...6 1 3 5 2 4 6 30 42 2 60 30 42 2 60 GFCI GFI PE 61...
Page 62: ...7 7 1 3 8x 2 3x 4 8x 8 8 62...
Page 63: ...9 52 9 1 4 7 2 5 8 3 4 20 6 9 0 1 63...
Page 64: ...1 0 2 18 AWG 3 10 52 3 4 10 3 1 PE 2 N 3 L1 4 1 PE 2 24 3 24 64...
Page 65: ...10 SC 6773200 sc sc Smart Probe Adapter 9321000 Smart Probe Adapter 11 65...
Page 66: ...11 1 0 2 18 AWG 2x 63 NO COM NC 4 20 4 20 4 20 2 Modbus 49937 66...
Page 67: ...40007 40008 12 0 0 8 2 5 30 63 12 TTL 1 12 3 TTL 2 2 4 2 DIG6 67...
Page 68: ...MSDS SDS 133 13 3 68...
Page 69: ......