Remove the access plugs
Install cables and conduit through the electrical access ports. Refer to
sealing plugs by pushing them out from inside the enclosure to unlock the seal, and then remove
completely by pulling from the outside. Remove knockouts as necessary from the electrical access
plate with a hammer and screwdriver. To keep the enclosure rating, put a cover on all ports that are
not used.
Figure 7 Electrical access ports
1
Power in (power cord only), no ground plate. Do not
use for conduit.
3
Communication and network modules (8x)
2
Communication and network modules (3x)
4
Power in or out (conduit or power cord), ground
plate, communication and network modules (8x)
Remove the access cover
Remove the access cover to connect to the wiring terminals. Refer to
.
Figure 8 Access cover removal
Wiring connections overview
shows all of the possible wiring connections. Make sure to use the wire gauge that is
specified for the connection (refer to
12
English
Summary of Contents for 5500sc
Page 2: ...English 3 Fran ais 18 Espa ol 35 52 2...
Page 54: ...54...
Page 56: ...1 2 3 4 5 pH 1 2 56...
Page 57: ...1 1 5 9 2 6 SD 10 3 7 1 11 4 8 12 2 1 57...
Page 58: ...4 2000 6562 2000 3 58...
Page 59: ...3 1 2 4 4 1 2 5 9 2 6 10 1 3 2 7 4 1 8 59...
Page 60: ...6 5 6 6 1 4 1 5 1 3 5 2 4 60...
Page 61: ...6 1 3 5 2 4 6 30 42 2 60 30 42 2 60 GFCI GFI PE 61...
Page 62: ...7 7 1 3 8x 2 3x 4 8x 8 8 62...
Page 63: ...9 52 9 1 4 7 2 5 8 3 4 20 6 9 0 1 63...
Page 64: ...1 0 2 18 AWG 3 10 52 3 4 10 3 1 PE 2 N 3 L1 4 1 PE 2 24 3 24 64...
Page 65: ...10 SC 6773200 sc sc Smart Probe Adapter 9321000 Smart Probe Adapter 11 65...
Page 66: ...11 1 0 2 18 AWG 2x 63 NO COM NC 4 20 4 20 4 20 2 Modbus 49937 66...
Page 67: ...40007 40008 12 0 0 8 2 5 30 63 12 TTL 1 12 3 TTL 2 2 4 2 DIG6 67...
Page 68: ...MSDS SDS 133 13 3 68...
Page 69: ......