![Hach 5500sc PO4 3-HR Operations Download Page 9](http://html1.mh-extra.com/html/hach/5500sc-po4-3-hr/5500sc-po4-3-hr_operations_3147064009.webp)
1.
Select GRAB SAMPLE>GRAB SAMPLE OUT.
2.
Complete the instructions on the display.
3.
Put the grab sample out tube in a clean container. The grab sample tube is located in the lower
unit of the cabinet on the left side.
4.
Push down on the spigot (upper left corner of the lower compartment) for the grab sample tube to
dispense a grab sample.
Set up the system
The configuration settings can be changed in the SETUP SYSTEM menu or in the analyzer setup.
Refer to
on page 6.
1.
Select SETUP SYSTEM.
2.
Push
menu
and select SETUP SYSTEM.
3.
Select an option.
Option
Description
MEAS MODE
Changes the mode of the measurement cycle. Options: interval or continuous
(default). The continuous mode measures approximately every 9 minutes.
EDIT INTERVAL
(conditional)
Changes the interval time when MEAS MODE is set to interval. Options:
10–240 minutes (default = 15 minutes).
MEAS UNITS
Changes the measurement units that are shown on the display and in the data
log. Options: ppb (default), ppm, mg/L, µg/L.
SIGNAL AVERAGE
Selects the number of measurements used to calculate an average measurement
(1–5). This reduces variability in measurements (default = 1, no averaging).
EDIT ANALYZER
NAME
Changes the name that is shown on the top of the measurement screen
(maximum of 16 characters).
EDIT CHANNEL NAME
Changes the name of the sample source that is shown on the measurement
screen (maximum of 10 characters).
CONFIGURE
SEQUENCER (optional)
Starts or stops measurements for individual sample sources. Sets the
measurement order of the sample sources when there is more than one sample
source.
SAMPLE MISSING
Sets the operation that occurs if no sample is detected. Options: ON DELAY
(default) or OFF DELAY. ON DELAY: the instrument waits the remainder of the
measurement cycle, then goes to the next sample in the sequence. OFF DELAY:
the instrument waits 10 seconds, then goes to the next sample in the sequence.
SET DATE & TIME
Sets the analyzer time and date.
DISPLAY SETUP
Changes the language. Adjusts the order measurements are shown. Adjusts the
display contrast settings.
DISABLE REMINDERS
Stops scheduled maintenance alerts for individual components. Options: tubing,
stir bar, colorimeter cell, air filter, stir motor, air relief valve, pinch valve, air
compressor, reagent valves, sample valves, standard valves, colorimeter LED, fan
filter, air check valve.
MANAGE DEVICES
Installs or removes input modules. Refer to
on page 10 for
more information.
INSTRUMENT
INFORMATION
Shows the analyzer information. Refer to
on page 16.
CALCULATION
Set up variables, parameters, units and formulas for the analyzer. Refer to
on page 10.
SETUP OUTPUTS
Selects and configures the 4–20 mA setup, the relay setup and the error hold
mode. Refer to
on page 11 for more information.
English
9
Summary of Contents for 5500sc PO4 3-HR
Page 208: ...208 216 211 221 213 SD 224 213 224 214 1 1 1 6 2 7 3 8 4 9 5 Enter 2 3 208...
Page 209: ...2 1 6 2 7 3 8 4 9 5 SD SD 10 3 1 3 Rx Sx 1 2 4 0 100 1 209...
Page 210: ...45 75 10 45 50 75 10 50 2 6 1 2 3 ppb 48 SD 4 SD 210...
Page 211: ...4 SD 1 3 2 2 4 SD 1 1 1 2 3 4 211 4 1 2 3 4 2 211...
Page 212: ...5 6 7 0 14 bar 2 psi 0 28 bar 4 psi 55 mL 8 ENTER 9 10 224 216 216 1 enter enter 1 224 212...
Page 213: ...MSDS SDS 1 2 1 100 1 100 1 100 100 100 100 1 250 500 mL 2 213...
Page 214: ...3 4 5 5 5 1 2 3 4 211 1 2 3 9 10 240 15 ppb ppm mg L g L 1 5 1 16 10 214...
Page 215: ...10 216 223 216 4 20 mA 216 Modbus Profibus HART HACH55 215...
Page 218: ...PID 5 50 50 6 1 2 3 219 216 218...
Page 219: ...1 2 3 1 0 0 9 1 0 1 1 1 0 0 5 1 0 1 5 5 4 0 0 5 3 5 4 0 5 0 300 5 0 300 5 4 10 219...
Page 220: ...5 5 5 0 0 0 0 6 5 30 7 8 1 220...
Page 221: ...2 3 4 5 6 1 1 2 18 000 18 000 1 2 2 3 221...
Page 224: ...SD SD SD SD 2 GB 1 SD 4 211 2 SD SD SD 3 SD SD SD SD 1 SD 2 SD SD SD 3 4 SD 5 1 224...
Page 225: ...1 225 225 4 20 mA 1 2 3 ppb 4 1 2 ENTER 3 4 225...
Page 279: ...2 1 6 2 7 3 8 4 9 5 SD SD 10 3 1 3 Rx Sx 1 2 PROGNOSYS 4 PROGNOSYS PROGNOSYS 0 100 1 279...
Page 280: ...45 75 10 45 50 75 10 50 2 1 2 home 3 48 SD 4 SD 280...
Page 281: ...4 SD 1 3 2 2 4 SD 1 1 1 2 3 4 281 4 1 2 3 4 5 6 2 281...
Page 282: ...7 0 14 2 0 28 4 55 8 enter 9 10 294 286 286 1 enter enter 1 294 MSDS SDS 282...
Page 283: ...1 menu 2 1 100 1 100 1 100 100 100 100 1 250 500 2 3 4 5 5 5 283...
Page 284: ...1 2 3 4 281 1 2 menu 3 9 10 240 15 ppb ppm 1 5 1 16 10 10 284...
Page 285: ...285 292 286 4 20 286 Modbus Profibus HART HACH55 1 menu 2 285...
Page 287: ...4 20 1 menu 4 20 2 3 4 5 50 50 6 287...
Page 288: ...1 menu 2 3 288 286 1 menu 2 3 1 0 0 9 1 0 1 1 1 0 0 5 1 0 1 5 5 288...
Page 289: ...4 0 0 5 3 5 4 0 5 0 300 5 0 300 5 4 10 5 5 5 ONMAX 0 OFFMAX 0 ONMIN 0 min OFFMIN 0 min 6 289...
Page 290: ...5 30 7 8 1 2 3 4 5 6 1 1 menu 2 290...
Page 292: ...2 0 1023 511 3 A D REF A D A D 6 6 6 1 2 3 1 2 292...
Page 294: ...1 SD 2 SD SD SD 3 4 SD 5 1 1 cal 295 294 4 20 1 cal 2 3 ppb 4 294...
Page 295: ...1 cal 2 enter 3 4 295...
Page 370: ...4 SD 1 3 LED 2 2 4 SD 1 1 1 2 3 4 370 4 1 LANGUAGE 2 DATE FORMAT 3 DATE TIME 2 370...
Page 386: ...1 cal START MANUAL CAL 2 enter 3 YES NO 4 386...
Page 404: ......
Page 405: ......