DIET CALIBRATION

There are three types of diet calibration available,
depending on the type of product. 

Automatic (diet-a and T.A.-a products only)

Automatic calibration performs a calibration
based on three successive stable measurement
cycles at the start of a product run, and assumes
the product is 

in spec 

when the run starts. After

three successive stable measurements, the new
Diet Calibration Coefficient is calculated based on
the target value for the product.
If there is a large difference between the new
coefficient and the last calibration, a warning
message will be displayed when you press the
Product b utton, showing the percenta ge
difference between the two values. 

Manual (diet-m and T.A.-m products only)

A dialog box appears automatically. The last five
measurements are initially set to zero.

Once the measurements displayed indicate a
stable condition (at least three successive stable
measurement cycles), press 

OK

 to close the

dialog box and to calculate the new slope value
for the current product, which is based on the
value in the edit box.
The target measurement for the product is
automatically displayed in the edit box, as the
calibration value. This can be changed based on
any Lab Tests performed, as it is used in
calculating the calibration coefficient. 

Absolute (diet and T.A. products only)

Before doing a calibration, it is necessary to check
your water calibration coefficient first. This can be
done during the next rinse-down period. 
Make sure you have set the 

Water Calibration

Status

 to 

On

 and select water (product number 0)

as the product for measuring. As you start the
measurement process, a new dialog box appears
automatically.

The dialog box will initially display zeros in the last
five measurements column. These values are
then updated every 15 seconds, as new as
measurements are taken.
When the measurements indicate a stable
condition, press 

OK

 to close the dialog box and

apply the most recent reading (

n1

) as the new

water coefficient. This new coefficient will be
applied as the water correction factor for all diet
and T.A. products. 
Once the water calibration is complete, you can
now go through the product calibration process.
Select the 

Calibration

 button in the 

Products

List

 window for the required product. This will

display the density of the syrup (at 20°C) and the
mixing ratio (volume of syrup to volume of water),
which can 

only

 be modified if this is a diet product

(not applicable to T.A. products).

The current Slope value is displayed, along with
the Water calibration coefficient calculated during
the water calibration run described above. These
can be changed if required.
Press 

Calculate

, to compute a new slope value

for the product. Press 

OK

 to apply these new

parameters to future measurements of this
product.

Calibration - From PC

Summary of Contents for ORBISPHERE 3624

Page 1: ...Operator Quick Guide ORBISPHERE 3624 Revision F 03 10 2008...

Page 2: ...accidental modifications The Display has a two line liquid crystal display LCD with 16 characters per line A Light Button when pressed provides illumination to the LCD for three minutes Four Function...

Page 3: ...h active LINE which are activated via the Start Line A B command in the Configuration menu or by the A B buttons on the toolbar at the top left of the screen The Numerical Indicator Panel shows inform...

Page 4: ...utput From the General window these include the title bars color or black and white the data displayed in either or both tabular and graphical form Choose the styles that suit you best Select the Plot...

Page 5: ...ts the instrument sets an alarm SERIAL OUTPUT instrument only The instrument serial output is normally used to interface to the PC for analysis of measurement data via the ProBrix Plus PC Program CONT...

Page 6: ...roducts on both Line A and B Autosave This option allows you to select the time delay for automatically saving measurement data Set to None no Autosave required End of Run save only on completion or a...

Page 7: ...ts three calibration cycles The gas and temperature measurements are displayed along with the current calibration cycle number On completion of the three cycles the instrument then shows the sensitivi...

Page 8: ...e diet and T A products only Before doing a calibration it is necessary to check your water calibration coefficient first This can be done during the next rinse down period Make sure you have set the...

Page 9: ...ere agent System does not start No display or backlight Improper line voltage Check instrument voltage Select proper voltage 115V or 230V on power supply card Blown fuse Unplug instrument and replace...

Page 10: ...bble should evolve during the measurement ramp If solenoid leaks more than 1 bubble in 15 seconds return sensor to Orbisphere Grounding mismatch between Orbisphere instrument and electronic accessorie...

Page 11: ...VIOLATED Lmt Overflow On On At the end of a measurement cycle if V3end 2500mV or V3end 4450mV Check membrane 0 PURGE VOLTAGE DRIFTING Chk N A Off Off At the beginning of a measurement cycle if V3beg...

Page 12: ......

Reviews: