Page 6
Setting a LINE Configuration
Up to two filling lines may be displayed in the
program window. Each line will be connected to a
separate PC COM port, which will have been set
up during the installation process.
Access the LINE you wish to configure, by
selecting the
Line A
or
Line B
option. Modify
parameters for that LINE by selecting the
Change
button.
Products List
The
Products List
option allows you to set a
number of parameters for each product. There is
one product list per LINE, each with a maximum
of 9998 products. Note that product zero is
reserved for water measurement, and product
9999 is the default product used by the system
when the LINE is inactive. The default values for
both of these products are pre-set and cannot be
modified.
First select the LINE you wish to enter/modify
details for, by selecting either LINE A or LINE B in
the top left box.
Then select a product using one of the following
methods:
•
By typing the product number, if known,
in the upper right-hand box
•
By using the right-hand scroll bar
•
By using the Search button to find a
product name
Once a product is selected, you can now enter/
modify the default parameters. Note that
modifications are only activated the next time a
run is started for this product.
Oxygen and Temperature Values
S e t t h e u n i t s a n d l i m i t s f o r o x y g e n a n d
temperature displays and alarms. The values are
applicable to
all
products on both Line A
and
B.
Autosave
This option allows you to select the time delay for
automatically saving measurement data. Set to
None
(no Autosave required),
End of Run
(save
only on completion) or at a time interval of 1, 2, 5,
10, 30 or 60 minutes.
As the year is not part of the filename, ensure
these measurement files (
MMDDHHMM.cbo
) are
backed up and deleted at the end of each year.
Water Calibration Status
For Diet measurements only, it may be desirable
to perform a water density calibration periodically.
This can be conveniently done during a long
rinse-down period. To enable this calibration,
select this option, and turn the status to
On
.
Alarms Delay
System alarms are disabled each time a new run
is started or when the
Alarms ON/OFF
button is
s e l e c t e d . T he P r o Br i x P l u s P C Pro g r a m
automatically re-enables these after a delay
period of 5 minutes.
Selecting this option and entering a value of
between 0 and 30 will override the default to the
new value.
Sound Alarms
The ProBrix Plus PC Program can trigger audible
alarm messages. To receive these messages on
your PC, it must be equipped with speakers and
have a sound card installed.
Serial Output
You can send formatted measurement and
product data to other computers on your network.
Select the serial communications port linked to
the other computer. You cannot use the same
ports that have already been configured for LINE
A and/or LINE B.
For detailed information on the format of the data
elements transmitted, please press the
Help
button on this window and then select
Additional
serial output
help.
PC Configuration Menu