background image

Page 7

BAROMETRIC PRESSURE SENSOR

The instrument internal barometric pressure
sensor is calibrated at the factory, and normally
requires no further attention. However, you may
w i s h   t o   c a l i b r a t e   i t   a g a i n s t   y o u r   o w n
instrumentation, or simply check the instrument
for accuracy.

CO

2

 SENSOR

The sensor can be calibrated using either:

1)

A source of 100% pure CO

2

 gas at a 

known elevated pressure (

Partial 

Pressure

)

2)

A known concentration of CO

2

 gas at a 

known pressure (

Fraction

)

3)

A known concentration of dissolved 
CO

2

 (

Dissolved

)

The 

Partial Pressure

 method is generally

recommended when measuring at higher line
pressures, and requires a precise in-line pressure
gauge to perform. 
For the 

Partial Pressure

 and 

Fraction

 methods,

shut off the sample flow to the sampling module,
then open the front panel and remove the CO

2

sensor from its flow chamber. Do not disconnect
the cables or purge gas connections from the
sensor. Before calibrating, make sure that the grill
in the front of the CO

2

 sensor is clean and dry.

Insert the sensor in the flow chamber, and tighten
with its collar.
The 

Dissolved

 method requires a known

concentration of CO

2

 dissolved in liquid as a

reference sample, flowing in-line through the
sample line.
For all of the three methods described above,
expose the sensor to the gas and stabilize the
reading by operating in Measurement Mode for
about 30 minutes. After this time, press 

ESC

 and

select 

CALIBRATE

 from the main menu, followed

by 

SENSOR CO2

 and the calibration method.

Then, select your calibration measurement unit.
Enter the gas concentration of the calibration
medium. Press 

ENTER

 to start the calibration

process. The instrument starts three calibration

cycles. The gas and temperature measurements
are displayed along with the current calibration
cycle number. 
On completion of the three cycles, the instrument
then shows the sensitivity of the sensor as a
percentage of the sensitivity determined when
calibration was last performed.
This percentage must be between 50% and 150%
in order to calibrate. If the percentage is outside
the limits, you will see the 

CALIBRATION OUT

OF BOUNDS

 message. You will need to press

ESC

 to continue. Check that the membrane does

not need to be replaced and that no leaks are
evident. 
If the percentage is within the limits press 

ENTER

to accept and complete the calibration process
(the message 

CALIBRATION COMPLETE

 will

appear briefly) or 

ESC

 to abort.

O

2

 SENSOR

The 

In Air

 method places the O

2

 sensor in water-

saturated air, to provide a known oxygen
reference against which to calibrate.
Shut off the pump and turn off the sample flow to
the sampling module. Open its front panel and
remove the O

2

 sensor from its flow chamber. Do

not disconnect the sensor cable.
Dry the sensor thoroughly, before placing the
sensor storage cap under tap water. Shake off
any excess water, but leave a few drops inside
the cap. Then, loosely place the storage cap back
on the sensor, holding it in place by a few turns of
its collar.
Select 

CALIBRATE

 from the main menu followed

by 

O2 IN AIR

 to start the calibration. The process

is then similar to the CO

2

 sensor calibration.

The 

Direct

 method calibrates the oxygen sensor

against a liquid sample containing a known level
of dissolved O

2

, flowing through the sample line.

Select 

CALIBRATE

 from the main menu followed

by 

O2 DIRECT

 and the calibration units. Then

enter the gas concentration of the calibration
sample to start the calibration. The process is
then similar to the CO

2

 sensor calibration.

Calibration - From Instrument

Summary of Contents for ORBISPHERE 3624

Page 1: ...Operator Quick Guide ORBISPHERE 3624 Revision F 03 10 2008...

Page 2: ...accidental modifications The Display has a two line liquid crystal display LCD with 16 characters per line A Light Button when pressed provides illumination to the LCD for three minutes Four Function...

Page 3: ...h active LINE which are activated via the Start Line A B command in the Configuration menu or by the A B buttons on the toolbar at the top left of the screen The Numerical Indicator Panel shows inform...

Page 4: ...utput From the General window these include the title bars color or black and white the data displayed in either or both tabular and graphical form Choose the styles that suit you best Select the Plot...

Page 5: ...ts the instrument sets an alarm SERIAL OUTPUT instrument only The instrument serial output is normally used to interface to the PC for analysis of measurement data via the ProBrix Plus PC Program CONT...

Page 6: ...roducts on both Line A and B Autosave This option allows you to select the time delay for automatically saving measurement data Set to None no Autosave required End of Run save only on completion or a...

Page 7: ...ts three calibration cycles The gas and temperature measurements are displayed along with the current calibration cycle number On completion of the three cycles the instrument then shows the sensitivi...

Page 8: ...e diet and T A products only Before doing a calibration it is necessary to check your water calibration coefficient first This can be done during the next rinse down period Make sure you have set the...

Page 9: ...ere agent System does not start No display or backlight Improper line voltage Check instrument voltage Select proper voltage 115V or 230V on power supply card Blown fuse Unplug instrument and replace...

Page 10: ...bble should evolve during the measurement ramp If solenoid leaks more than 1 bubble in 15 seconds return sensor to Orbisphere Grounding mismatch between Orbisphere instrument and electronic accessorie...

Page 11: ...VIOLATED Lmt Overflow On On At the end of a measurement cycle if V3end 2500mV or V3end 4450mV Check membrane 0 PURGE VOLTAGE DRIFTING Chk N A Off Off At the beginning of a measurement cycle if V3beg...

Page 12: ......

Reviews: