12
Installation
3.2 Installation guidelines for the MET ONE HHPC–6
The MET ONE HHPC–6 is ready to use when shipped and has default configuration
settings (
section 4.4 on page 18
). Refer to the general guidelines during installation:
1.
Lift the MET ONE HHPC–6 from the shipping container or the carrying case and
place it on a flat surface.
2.
Remove the layer of protective plastic from the LCD screen
3.
Remove the isokinetic probe cover (black plastic cap) before the startup.
Note:
The pump and other internal electrical components will be damaged if the cap is left on.
3.3 Mechanical installation
The MET ONE HHPC–6 is shipped with two different intake devices:
•
Isokinetic probe
•
Intake hose barb
Important Note:
Do not use a wrench to connect or disconnect the intake nozzle or
isokinetic probe. Finger-tighten the connection.
3.3.1 Install the isokinetic probe
The isokinetic probe (refer to
Figure 4 on page 13
) is used to sample ambient air
samples. The isokinetic probe is pre-installed and ready for use as soon as the MET ONE
HHPC–6 is unpacked. To install the isokinetic probe:
1.
Attach the isokinetic probe to the intake nozzle on the MET ONE HHPC–6 by turning
the isokinetic probe to the right.
2.
Remove the black plastic cap attached to the isokinetic probe.
Note:
Make sure that the cap is attached to the isokinetic probe when the probe is not in use.
The cap protects the MET ONE HHPC–6 from debris and contamination.
To sample areas that are difficult to reach, connect longer tubing to the isokinetic probe
with a hose barb end.
Note:
Use a maximum of four feet of high purity or approved tubing only. Incorrect size or type of
tubing will affect the sample accuracy.
3.3.2 Install the intake hose barb
Connect the intake hose barb (refer to
Figure 4 on page 13
) to the MET ONE HHPC–6
for zero counting and purging or to sample areas that are difficult to reach.
To install the intake hose barb:
1.
Unscrew the isokinetic probe from the MET ONE HHPC–6.
2.
Attach the hose barb to the MET ONE HHPC–6.
3.
Remove the protective cap of the intake hose barb.
Summary of Contents for MET ONE HHPC-6
Page 1: ...701229 MET ONE HHPC 6 Handheld Airborne Particle Counter USER MANUAL May 2008 Edition 4 ...
Page 3: ......
Page 7: ...6 Specifications ...
Page 10: ...9 General Information Figure 2 MET ONE HHPC 6 1 Front view 3 Back view 2 Bottom view ...
Page 11: ...10 General Information ...
Page 35: ...34 Maintenance ...
Page 39: ...38 Replacement Parts and Accessories ...
Page 42: ...41 Section 10 Certification ...
Page 43: ...42 Certification ...
Page 45: ...44 ...
Page 47: ......