11
Housing version:
Figure shows HST-M2 Ident.-No.: 10342
Assembly group
A
Lid screws
B
Housing lid
C
Housing box
D
Wall mounting
Open the housing box [C] by unscrewing the 4 outer lid screws [A] of the housing box [C] and
lift the housing lid [B] from the housing box [C] with the HST-M. Make sure the protective
earthing conductors are not damaged. Securing the housing lid [B] against falling down.
Mount the enclosed wall mounting [D] according to the illustration.
Attach the housing box [C] with suitable screws and nuts if necessary.
Secure the screw connection against coming loose.
Taking the conditions on site into account, create a hole to thread the cables into the housing
box [B] insert a suitable cable gland with a protection class of at least IP 65 (
not
included in
scope of delivery).
Carry out the electrical connection (cf. section 12.4).
Place the housing lid [B] back on the housing box [C] with the HST-M2 and screw it in place
with the 4 lid screws [A]. Make sure the protective earthing conductors are not squashed or
disconnected.
Attention!
Make sure the protective earthing conduc-
tor is not squashed or disconnected.
No liability is accepted in the event of improper installation!
A
B
C
D