PPH-1503 User Manual
46
time is manully set. The automatic
Integration time can automatically
detect pulses in the 80uS to 833mS
range.
•
The manual time range setting is
33uS to 833333uS. The default
units are in microseconds (uS). The
unit will automatically round
down the last two digits to 00, 33
or 66 micoseconds. For example,
for a value of 65.999, it will be
rounded down to 33 and 66.01 will
be rounded down to 66.
TrigDelay
•
Trigger Delay
•
When a pulse is detected, there
will be a 25us code execution
delay time. The trigger delay
settings are used to filter out the
current overshoot. Measurement
will begin from after the trigger
delay time. The trigger delay
setting range is: 0~0.10000S, with a
resolution of 0.00001S. The setting
units are in seconds.
Note
The trigger delay setting range for Pulse
Current Digitization is 0 ~ 5S. See page
113.