REMOTE CONTROL
113
Example
:SENSe:PCURrent:TIME:AVERage?
Returns the integration time for the average
measurement.
Note
When in the pulse current digitization mode, the
IntTime setting is automatically changed to 33µs.
Command
:SENSe[1]:PCURrent:SYNChronize[:STATe] <b>
Function
Sets the triggering option for pulse current
measurement.
Description
<b>
0 /OFF: Digital trigger mode.
1/ON: Pulse level trigger mode.
Example
:SENSe:PCURrent:SYNChronize ON
The trigger mode is set to the pulse level trigger.
Note
Pulse Current Digitization:
When using remote control to read data, first set the
number of reading counts for the batch between
1~5000. See the command:
SENSe[1]:PCURrent:AVERage <NRf>
Command
:SENSe[1]:PCURrent:SYNChronize[:STATe]?
Function
Queries the pulse current measurement triggering
option.
Example
:SENSe:PCURrent:SYNChronize?
Returns the pulse current trigger option.
Command
:SENSe[1]:PCURrent:SYNChronize:DELay <NRf>
Function
Sets the trigger delay time.
Description
<NRf>
0~0.1 or 0~5 (Pulse current digitization)
北京海洋兴业科技股份有限公司(证券代码:839145) 电话:010-62176775 网址:www.hyxyyq.com