REMOTE
CONTROL
169
Preset
con:preset Presets
the
GSP-830.
System
con:sys:setup:save
Save the current system setting. Parameter: 1to10
Example: con:sys:setup:save_1 (save current setup to setup1)
con:sys:setup:recall
Recalls a system setting from setup file. Parameter: 1to10
Example: con:sys:setup:recall_1 (recall setting in setup1 file)
con:sys:gpibaddr?
Returns the current GPIB address.
con:sys:gpibaddr
Sets the GPIB address. Example: con:sys:gpib:addr_2
con:sys:auxsig
Turns On/Off the auxiliary signal. Parameter: on, off
Example: con:sys:auxsig_on (auxiliary signal On)
con:sys:clock:date?
Returns the current date setting.
Parameter: year/month/ day/ day of week 1 (Mon) to 7(Sun)
Example: 2006 6 24 6 (June 24
th
, Saturday, 2006)
con:sys:clock:date
Sets the date.
Parameter: year/month/day/ day of week 1 (Mon) to 7(Sun)
Example: con:sys:clock:date_2006_6_24_6 (Jun24, Sat, 2006)
con:sys:clock:time?
Returns the current time setting.
Parameter: hour / minute / second
Example: 13 30 26 (1p.m., 30 minutes, 26 second)
con:sys:clock:time
Sets the time. Parameter: hour / minute / second
Example: con:sys:clock:time_13_30_26 (1p.m., 30min, 26sec)
con:sys:clock:show
Turns On/Off clock display. Parameter: on, off
Example: con:sys:clock:show_on (clock display On)
con:sys:selftest?
Returns the self test result. Parameter: 0 (fail), 1 (pass) in the
following order: GPIB/Flash/SDRAM/RTC
Example: 1 1 0 1 (GPIBpass,Flashpass,SDRAMfail,RTCpass)
con:sys:lang
Selects language. Parameter: 1 (English), 2 (Simplified
Chinese) Example: con:sys:lang_2 (simplified Chinese)
con:sys:ser?
Returns the serial number. Example: EE8300000
con:sys:swver?
Returns the software version.
Example: v1.0 06/07/28 (version 1.0, 2006 July 28
th
)
con:sys:fwver?
Returns the firmware version. Example: v1.0 (version 1.0)
con:sys:hwver?
Returns hardware version in following order. RF, IF, DSP, MB
Example: v1.0 v1.0 v1.01 v1.0 (RF: version 1.0, IF: version 1.0,
DSP: version 1.01, MB: version 1.0)
con:sys:optstatus?
Returns optional items installation status in the following
order. 300HzRBW, EMIFilter, 10k/100kHzRBW, TG,
Demodulator, Medref (±1ppm stability) Parameter: 0 (not
installed), 1 (installed)
Example: 0 0 1 1 1 (TG, Demodulator, Medref are installed)