GSP-730 User Manual
96
meas:tra:read
Set
Query
Description
Returns the all the trace data for the selected trace.
Query syntax
meas:tra:read? <trace>
Parameter
<trace>
1
Trace A
2
Trace B
3
Trace C
all
All traces
Return parameter <trace
data>
Comma separated data values
encapsulated in brackets. i.e.,
{-92, -91, -90, ………-81}
Example
meas:tra:read? 1
>{ -92, -91, -90, -90, -90, -88, ……., -89, -92, -92, -91 }
Returns the trace data for the selected trace(s). A total
of 501 trace points are returned, from the start
frequency to the stop frequency. If “all” is selected, the
trace data is returned in three lots, {trace A}{trace
B}{traceC}. The units are in decibels. If the selected
trace is not active, 0s will be returned for each trace
point.
Power Measurement Commands
meas:acpr .................................................................................. 96
meas:acpr:lower? ...................................................................... 97
meas:acpr:upper? ..................................................................... 97
meas:ocbw ................................................................................ 97
meas:ocbw:bw? ........................................................................ 98
meas:ocbw:chpw? .................................................................... 98
meas:acpr
Set
Query
Description
Turns the ACPR function on or off, or queries its
status.
Syntax
meas:acpr {on|off}