BASIC OPERATION
39
Activating a Marker
There are two basic marker types, normal markers and delta
markers. Normal markers are used to measure the frequency/time
or amplitude of a point on the trace. Delta markers are used to
measure the difference between a reference point and a selected
point on the trace.
Activate a Normal Marker
Operation
1.
Press
Marker
>
Marker[F1]
and select a marker
number.
Marker:
1~5
2.
Press
[F2]
to turn the selected marker on.
3.
Press
Mode[F3]
and set the marker mode to
Normal.
4.
The display will show the marker on the trace
(centered by default) with the marker
measurement at the top of the display.
Maker No., Frequency, Amplitude
Marker