![GW Instek GSP-730 User Manual Download Page 53](http://html1.mh-extra.com/html/gw-instek/gsp-730/gsp-730_user-manual_2280065053.webp)
BASIC OPERATION
51
Adjacent channel offset
1 ~ 2
The frequency distance
between the adjacent
channels and main
channel.
Range: 1
Between 0Hz~3GHz
(0Hz excepted)
OCBW%
The ratio of occupied
bandwidth to the
amount of power
consumed.
Range: 0% to 100%, 0.1%
resolution.
ACPR
Description
Adjacent channel power refers to the amount of
power leaked to the adjacent channel from the
main channel. This measurement is a ratio of
the main channel power to power in the
adjacent channel.
Example
Main
CHBW
ADJ
CH1
ADJ
CH2
ADJ
CH1
ADJ
CH2
Offset 1
Offset 2
Channel spacing
To next main
channel
Operation:
Setting up the
main channel
1.
Press
Meas
>
ACPR[F2]
and turn ACPR on.
Any other measurement mode will automatically be