![GW Instek GPM-8213 User Manual Download Page 92](http://html1.mh-extra.com/html/gw-instek/gpm-8213/gpm-8213_user-manual_2280057092.webp)
GPM-8213 User Manual
92
No items (
“
-----
”
)
ASCII: NAN (Not A Number)
Error Data
Data does not exist (the display shows “-----”)
ASCII: NAN (Not A Number)
:NUMeric[:NORMal]:NUMBer
Description
Sets or returns the specified numeric data output
item function.
Syntax
:NUMeric[:NORMal]:ITEM<x> {<Function>}ALL}
Query Syntax
:NUMeric[:NORMal]:NUMBer?
Parameter/
Return parameter
<NRf>
1 to 34(ALL)
Example
:NUMERIC:NORMAL:NUMBER 10
:NUMERIC:NORMAL:NUMBER
->:NUMERIC:NORMAL:NUMBER 10
Note
If the parameter is omitted from
the :NUMeric[:NORMal]:VALue? command, the
numeric data items from 1 to the specified value
are output in order.
By default, the number of numeric data items is set
to 3.
:NUMeric[:NORMal]:ITEM<x>
Description
Sets or returns the specified numeric data output
item function.
Syntax
:NUMeric[:NORMal]:ITEM<x> {<Function>}
Query Syntax
:NUMeric[:NORMal]:ITEM<x>?
Set
Query
Set
Query