![GW Instek GPM-8213 User Manual Download Page 37](http://html1.mh-extra.com/html/gw-instek/gpm-8213/gpm-8213_user-manual_2280057037.webp)
BASIC SETTING
37
Setting up auto-zero function
Steps
1.
Press
Setup
button.
Setup
2.
Press
Enter
button.
Enter
3.
Press down arrow key four times.
x4
4.
Press
Enter
button to enter
Auto
Zero
item. Use up and down arrow
keys to select the desired option
and then press
Enter
button again
to confirm your selection.
Enter
Enter
Option
On: Auto-zero function is activated once per
hour or when range is switched
Off: Auto-zero function is only activated once
when the range is switched. The auto-zero
function is turned off when the integrator
function is executed
Default value
Off