COMMAND OVERVIEW
83
Example
:INPUT:VOLTAGE:AUTO ON
:INPUT:VOLTAGE:AUTO?
->:INPUT:VOLTAGE:AUTO 1
[:INPut]:CURRent:RANGe
Description
Sets or returns the current range.
Query
[:INPut]:CURRent:RANGe {<Current>}
Query Syntax
[:INPut]:CURRent:RANGe?
Parameter/
Return parameter
<Current>
5, 10, 20, 50, 100, 200, 500(mA)
1, 2, 5, 10, 20(A) when the crest factor is
set to 3.
2.5, 5, 10, 25, 50, 100, 250(mA)
0.5, 1, 2.5, 5, 10(A)when the crest factor is
set to 6
Example
:INPUT:CURRENT:RANGE 20A
:INPUT:CURRENT:RANGE?
->:INPUT:CURRENT:RANGE 20.0E+00
[:INPut]:CURRent:AUTO
Description
Sets or returns the current auto range on/off state.
Syntax
[:INPut]:CURRent:AUTO {<Boolean>}
Query Syntax
[:INPut]:CURRent:AUTO?
Parameter
<Boolean>0 OFF
<Boolean>1 ON
Return parameter 0
Turn the current auto range function off.
1
Turn the current auto range function on.
Example
:INPUT:CURRENT:AUTO ON
:INPUT:CURRENT:AUTO?
->:INPUT:CURRENT:AUTO 1
Set
Query
Set
Query