触发指令
READ
Query
Description
Returns the measurement value.
Query Syntax
READ?
Return parameter
<NR3>
±0.0000~5.1000E±X
Example
READ?
>+2.2012E+0
Returns the measurement.
Note
+9.9999E+9 : indicate the HVP is detected.
MEASure<X>
Query
Description
Returns the results of the selected channel in the scan
mode, including HI/LO/IN and value.
Query Syntax
MEASure<X>?
Parameter
<X>
Channel 1~100
Return parameter
0|1|2,<NR3>
0:LO
1:IN
2:HI
<NR3>: Measurement result.
Example
MEAS1?
>1,+0.9978E+1
Returns channel 1 as 9.978Ω.