GOM-804 & GOM-805
使用手册
1
3
8
TEMPerature:AMBient:DATa
Set
Query
Description
Sets or returns the user-set ambient temperature value
for the temperature compensation and the temperature
conversion function.
Syntax
Query Syntax
TEMPerature:AMBient:DATa <NRf>
TEMPerature:AMBient:DATa?
Parameter
<NRf>
-50.0~399.9 (Unit: º
C)
Return parameter
<NR2>
-50.0~399.9 (Unit: º
C)
Example
TEMP:AMB:DAT 25.6
Sets the user ambient temperature value to +25.6º
C.
TEMP:AMB:DAT?
>25.6
Returns the set ambient temperature as 25.6º
C.
TEMPerature:AMBient:STATe
Set
Query
Description
Sets or returns the state of the user-set ambient
temperature.
Syntax
Query Syntax
TEMPerature:AMBient:STATe <NR1> | {OFF|ON}
TEMPerature:AMBient:STATe?
Parameter/
Return parameter
<NR1>
0:OFF.
1:ON.
OFF
Disables the user-set ambient
temperature.
ON
Enables the user-set ambient
temperature.
Example
TEMP:AMB:STAT OFF
Disables the user-set ambient temperature.
TEMPerature:COMPensate:COEFficient
Set
Query
Description
Sets or returns the temperature coefficient for
temperature compensation function.
Syntax
Query Syntax
TEMPerature:COMPensate:COEFficient <NR1>
TEMPerature:COMPensate:COEFficient?
Summary of Contents for GOM-804
Page 1: ...直流微欧姆计 GOM 804 GOM 805 使用手册 固纬料号 ISO 9001 认证企业 ...
Page 21: ...GOM 804 GOM 805 使用手册 2 1 设置 倾斜站立 倾斜 将支脚向前扳 如下所示 直立 将支脚向后扳 如下所示 ...
Page 38: ...GOM 804 GOM 805 使用手册 38 Internal trigger source ...
Page 57: ...GOM 804 GOM 805 使用手册 57 GUARD SOURCE SENSE SENSE SOURCE shielding ...
Page 96: ...GOM 804 GOM 805 使用手册 9 6 功能检测完成 ...
Page 137: ...GOM 804 GOM 805 使用手册 1 37 温度指令 ...
Page 142: ...GOM 804 GOM 805 使用手册 1 42 Example TEMP UNIT DEGF Sets temperature unit to º F Fahrenheit ...