GOM-804 & GOM-805
使用手册
1 2 0
Example
CALC:COMP:PERC:UPP 90.00
Sets the upper limit percent value to +90.00%.
CALC:COMP:PERC:UPP?
>90.00
Returns the upper limit as +90.00%.
CALCulate:SCAN:CHANnel
Set
Query
Description
Sets or returns the channel for the scan function.
Syntax
Query Syntax
CALCulate:SCAN:CHANnel <NR1>
CALCulate:SCAN:CHANnel?
Parameter/
Return parameter
<NR1>
1~100
Example
CALC:SCAN:CHAN 5
Sets the channel to 5.
CALCulate:SCAN:DELay
Set
Query
Description
Sets or returns the interval delay for the scan function.
Syntax
Query Syntax
CALCulate:SCAN:DELay <NR1>
CALCulate:SCAN:DELay?
Parameter/
Return parameter
<NR1>
400~30000
Unit:ms
Example
CALC:SCAN:DEL 500
Sets interval delay of the scan to 500ms.
CALCulate:SCAN:LIMit:LOWer
Set
Query
Description
Sets or returns the lower limit value for the scan
function.
Syntax
Query Syntax
CALCulate:SCAN:LIMit:LOWer {<NRf>[,<String>]}
CALCulate:SCAN:LIMit:LOWer?
Parameter
<NRf>
000.0000~999.9999
<String>
mohm/ohm/kohm/maohm,unit
If unit is not set, the unit will be
automatically set by the present
range.
Summary of Contents for GOM-804
Page 1: ...直流微欧姆计 GOM 804 GOM 805 使用手册 固纬料号 ISO 9001 认证企业 ...
Page 21: ...GOM 804 GOM 805 使用手册 2 1 设置 倾斜站立 倾斜 将支脚向前扳 如下所示 直立 将支脚向后扳 如下所示 ...
Page 38: ...GOM 804 GOM 805 使用手册 38 Internal trigger source ...
Page 57: ...GOM 804 GOM 805 使用手册 57 GUARD SOURCE SENSE SENSE SOURCE shielding ...
Page 96: ...GOM 804 GOM 805 使用手册 9 6 功能检测完成 ...
Page 137: ...GOM 804 GOM 805 使用手册 1 37 温度指令 ...
Page 142: ...GOM 804 GOM 805 使用手册 1 42 Example TEMP UNIT DEGF Sets temperature unit to º F Fahrenheit ...