CALCulate:COMPare:LIMit:MODE
Set
Query
Description
Sets or returns the compare mode for the compare function.
Syntax
Query Syntax
CALCulate:COMPare:LIMit:MODE {ABS|DPER|PER}
CALCulate:COMPare:LIMit:MODE?
Parameter/
Return parameter
ABS
The test results are judged from absolute
values.
DPER
The test results are judged from a
reference value ± a percentage offset.
(delta percentage)
PER
The test results are displayed as a
percentage of the reference value.
Example
CALC:COMP:LIM:MODE ABS
Sets test results as absolute values for the compare function.
CALCulate:COMPare:LIMit:LOWer
Set
Query
Description
Sets or returns the lower limit value for the compare
function.
Syntax
Query Syntax
CALCulate:COMPare:LIMit:LOWer {<NRf>[,<String>]}
CALCulate:COMPare:LIMit:LOWer?
Parameter
<NRf>
000.0000~999.9999
<String>
mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present range.
Return parameter
<NR3>
000.0000~999.9999E±X
Example
CALC:COMP:LIM:LOW 0.95,kohm
Sets the lower limit value to 0.95kΩ.
CALC:COMP:LIM:LOW?
>0.9500E+3
Returns the lower limit as 0.95kΩ.
Note
This command is only applicable when compare mode is
set to ABS for compare function.
Summary of Contents for GOM-804
Page 1: ...直流微欧姆计 GOM 804 GOM 805 使用手册 固纬料号 ISO 9001 认证企业 ...
Page 21: ...GOM 804 GOM 805 使用手册 2 1 设置 倾斜站立 倾斜 将支脚向前扳 如下所示 直立 将支脚向后扳 如下所示 ...
Page 38: ...GOM 804 GOM 805 使用手册 38 Internal trigger source ...
Page 57: ...GOM 804 GOM 805 使用手册 57 GUARD SOURCE SENSE SENSE SOURCE shielding ...
Page 96: ...GOM 804 GOM 805 使用手册 9 6 功能检测完成 ...
Page 137: ...GOM 804 GOM 805 使用手册 1 37 温度指令 ...
Page 142: ...GOM 804 GOM 805 使用手册 1 42 Example TEMP UNIT DEGF Sets temperature unit to º F Fahrenheit ...