REMOTE CONTROL
97
Example
:FUNC:CSIN:SDIP CLP1,100,SQU,50,50
Sets the surge/dip arbitrary waveform parameters as
site=square, ACV=50%, site=50%.
[:SOURce]:FUNCtion:CSINe:STAircase
Set
Query
Description
Sets or queries the staircase waveform type and the
number of “steps” in the waveform.
Note:
The :SOURce:FUNCtion:CSINe:TYPE command
must first be used to set the save slot number
(CLP1|2|3) and the waveform type (STAircase)
before this command can be executed.
Syntax
[:SOURce]:FUNCtion:CSINe:STAircase
{CLP1|CLP2|CLP3,<NR2>|MINimum|MAXimum}
Query Syntax
[:SOURce]:FUNCtion:CSINe:STAircase?
{CLP1|CLP2|CLP3[,MINimum|MAXimum]}
Parameter/Return
parameter
CLP1
Save slot 1
CLP2
Save slot 2
CLP3
Save slot 3
<NR2>
1 ~100 steps
MINimum
1 step
MAXimum
100 steps
Example
:FUNC:CSIN:STA CLP1,50
Sets the staircase arbitrary waveform at memory CLP1
to have 50 steps.
[:SOURce]:FUNCtion:CSINe:TYPE
Set
Query
Description
Sets the selected save slot to a type of arbitrary
waveform or queries its state.
Note:
This command must be used before one of the
[:SOURce]:FUNCtion:CSINe:CFACtor,
[:SOURce]:FUNCtion:CSINe:CLIP,
[:SOURce]:FUNCtion:CSINe:SDIP or
[:SOURce]:FUNCtion:CSINe:STAircase commands
are used.