2.04
Safety functions
The emergency stop and emergency lower should only be used in the event
of hoist failure or fault. The fault must be identified and rectified by a suitably
qualified technician before the hoist can be taken back into use. The Emer
-
gency stop should not be reset unless by a suitably qualified technician.
Under no circumstances should the user attempt to reset and continue to use
the hoist where the emergency stop has been activated due to fault or failure.
The hoist issue should be referred for service and rectification by a suitably
qualified technician.
Emergency stop and lowering strap
The red strap has the following functions:
• One pull: Emergency stop is activated.
• Constant pull: Emergency lowering is activated.
Emergency stop
If the GHZ does not stop/react to the hand control when the GHZ is in
use, pull the red strap and the lifting/lowering functions (except emergency
lowering) are deactivated. When the emergency stop is activated, the hoist
will not function. The green lamp is switched off.