Service and lifetime
5.01
Lifetime
The products have an expected lifetime of 15 years, on the condition of
correct use and correct service inspections, see section 5.02.
Replacement of components
Replacement of batteries, PCBs and lifting straps must be performed by a
qualified service technician or the Guldmann Service Team.
No part of the equipment shall be serviced when in use with a patient.
5.02
Safety/service inspections
In accordance with international standard EN/ISO 10535 “Hoist for the
transfer of disabled persons – Requirements and test methods” an inspection
should be performed on the hoist at least once a year.
Guldmann recommends that regular safety/service inspection is performed at
least once a year with regard to the pattern of usage.
Safety/service inspections of the products must be performed by a qualified
service technician or the Guldmann Service Team. In connection with the
purchase Guldmann may offer a service agreement for this inspection.
During the safety/service inspection a report must be prepared on what was
checked and replaced. Parts that are worn or defective must be replaced with
new parts from Guldmann. Spare parts drawings and documentation can be
obtained from the manufacturer or supplier.
Documentation/checklist regarding safety/service inspection can be
obtained from the manufacturer or supplier.
5.03 Troubleshooting
The hoist does not respond to the hand control’s keys
1. Check that the emergency stop is not activated
(see section 2.04)
.
2. Check that the hoist has power supply and that the battery is recharged
(see section 2.02)
.
3. Check that the power supply is switched on and connected to the charging
station.
4. Place the hand control in the charging station and recharge the hoist
(see section 2.03)
.
5. Contact the Guldmann Service Team if the fault cannot be found and
corrected.