7
designed and supplied by other manufacturers.
8.3 The accessory of the product should be cleaned by clean water or
sterilized liquid. Do not soak.
8.4 Please clean the resin remained on the top of the main unit after
using to avoid infecting the life-span or solidified effect.
9. Trouble shooting
Faulty
Possible cause
Solutions
No indication
No response.
1. Battery is out of
power.
2. Battery is protected.
3. Faulty of battery.
1. Charging.
2. Please put the curing
light into the pedestal for
charging, then the battery
works again.
3. Please contact our
special repair shop or us.
Light intensity is
weak.
There is resin on the
top of the optical fiber.
1. Clean the resin.
2. Change a new optical
fiber.
T h e e q u i p m e n t
d o e s n ’ t c h a r g e
when the adapter is
connected.
1. The adapter is not
connected well
2. The charging point
is impurity.
3. Faulty of adapter or
incompatible.
1. Reconnect.
2. Clean by the alcohol.
3. Please contact our
special repair shop or us.
Effective duration
o f t h e b a t t e r y
become short.
The capacity of the
battery decreased.
Please contact our special
repair shop or us.
The OLED screen
t w i n k l e s w h e n
charging.
low voltage.
Back to normal after 15
minuets charging.
If all the above solutions have been completed, the machine still can not
work normally. Please contact our special repair shop or us.