11
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.
Guidance & Declaration - Electromagnetic immunity
The models X-Cure are intended for use in the electromagnetic environment specified
below. The customer or the user of the models X-Cure should assure that it is used in
such an environment.
I m m u n i t y
test
IEC 60601 test
level
C o m p l i a n c e
level
Electromagnetic environment -
guidance
C o n d u c t e d
R F l E C
61000-4-6
Radiated RF
lEC 61000-4-
3
3 Vrms
150 kHz to 80 MHz
6 Vr m s i n I S M
bands
3 V/m
80 MHz to 2.7 GHz
385MHz-5785MHz
Test specifications
for ENCLOSURE
PORT IMMUNITY
t o R F w i r e l e s s
c o m m u n i c a t i o n
equipment
(Refer to table 9
of IEC 60601-1-
2:2014)
3 Vrms
150 kHz to 80
MHz
6 Vrms in ISM
bands
3 V/m
80 MHz to 2.7
GHz
3 8 5 M H z -
5785MHz Test
specification s
for ENCLOSUR
E P O R T
IMMUNITY to
R F w i r e l e s s
communicati on
equipment
(Refer to table 9
of IEC 60601-1-
2:2014)
P o r t a b l e a n d m o b i l e R F
communications equipment
should be used no closer to
any part of the models X-Cure,
including cables, than the
r e c o m m e n d e d s e p a r a t i o n
distance calculated from the
equation applicable to the
frequency of the transmitter.
R e c o m m e n d e d s e p a r a t i o n
distance
d=[3,5/V1]×P
1/2
d=1.2×P
1/2
80 MHz to 800 MHz
d=2.3×P
1/2
800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter
In watts (W) according to the
transmitter manufacturer and d
Is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,
a
should be less than
the compliance level in each
frequency range.
b
Interference may occur In the
vicinity of equipment marked with
the following symbol:
NOTE I At 80 MHz end 800 MHz. the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.