5
immediately by short pressing the button. If the main unit gets hot
obviously, please turn off the device until the main unit becomes cool.
Please don’t make it continuously illuminate more than 10 times.
5.9 Low power detect circuit is fixed inside of the main unit, when low
power is detected, the battery symbol on OLED screen become wink,
please charge in time.
5.10 Connect the adapter well when charge. Inserted the main unit into
the charging seat and pressing out the buckle between the main unit.
Then the OLED screen will scroll. If detect battery damage or other
abnormal, the charging indicator lights flashing.
5.11 When operating finish, please clean the optical fiber with calico to
avoid infecting the light intensity.
5.12 This equipment will turn off automatically if don't any action within
2 minutes, turn it on by press power button.
5.13 The effective light intensity of this equipment is much more higher
than Halogen Lamp, The solidified depth of the curing light composites
resin for 10 seconds will not less than 4mm.
6. Precaution
6.1 The optical fiber should be autoclaved under the high temperature of
134℃ and pressure 0.22 Mpa.
6.2 Please recharge the battery at least 4 hours before first time usage.
6.3 During operation, the light should be aimed straightly at the resin to
ensure the effect of solidification.
6.4 Be sure to use the original light hood to avoid the blue light hurt
eyes. Prohibit aiming light at eyes directly.
6.5 Only the original adapter could be used, because other brand adapters
are likely to damage the circuit.
6.6It is forbidden to use metal or other conductors to touch the charging
point of main unit, because it may burn the internal circuit or make the
lithium short circuit.
6.7 Charging the battery in the condition of cool and ventilated. Please
make sure of pressing out the buckle between the main unit and the