Guidance & Declaration — electromagnetic immunity 

The models X-Cure are intended for use in the electromagnetic environment specified 

below. The customer or the user of the models X-Cure should assure that It is used in 

such an environment. 

Immunity test

IEC 60601 test 

level 

Compliance level

Electromagnetic 

environment - 

guidance 

E l e c t r o s t a t i c 

discharge (ESD)  

lEC 61000-4-2 

±8 kV contact 

±2 kV, ±4 kV, ±8 

kV, ±15 kV air

±8 kV contact ±2 kV, 

±4 kV, ±8 kV, ±15 kV 

air

F l o o r s   s h o u l d   b e 

wood, concrete or 

ceramic tile. If floors 

a r e   c o v e r e d   w i t h 

synthetic material, 

the relative humidity 

should be at least 30 

%. 

E l e c t r i c a l   f a s t 

transient/burst  

IEC 61000-4-4

±2kV for power 

supply lines 

±1 kV for Input/

output lines

± 2 k V   f o r   p o w e r 

supply lines ±1kV 

for interconnecting 

cable

Mains power quality 

should be that of a 

typical commercial or 

hospital environment. 

Surge  

lEC 61000-4-5

±1 kV line to line 

±2 kV line to earth

±1 kV line to line

Mains power quality 

should be that of a 

typical commercial or 

hospital environment.

Voltage dips, short 

interruptions and 

voltage variations on 

power supply input 

lines  

IEC 61000-4-11.

<5 % UT 

(>95% dip in UT.) 

for 0.5 cycle  

40 % UT  

(60% dip in UT)  

for 5 cycles  

70% UT  

(30% dip in UT)  

for 25 cycles  

<5% UT 

(>95 % dip in UT)  

for 5 sec

<5 % UT 

(>95% dip in UT.) 

for 0.5 cycle  

40 % UT  

(60% dip in UT)  for 

5 cycles  

70% UT  

(30% dip in UT)  for 

25 cycles  

<5% UT 

(>95 % dip in UT)  

for 5 sec

Mains power quality 

should be that of a 

typical commercial or 

hospital environment. 

I f   t h e   u s e r   o f   t h e 

models X-Cure require 

continued operation 

during power mains 

interruptions, it is 

recommended that 

the models X-Cure 

be powered from an 

uninterruptible power 

supply or a battery. 

Power frequency 

(50/60 Hz) magnetic 

field  

lEC 61000-4-8  

30A/m

30A/m

P o w e r   f r e q u e n c y 

m a g n e t i c   f i e l d s 

should be at levels 

characteristic of a 

typical location in a 

typical commercial or 

hospital environment. 

Summary of Contents for X-Cure

Page 1: ......

Page 2: ...nting 4 5 Operation 4 6 Precaution 5 7 Contraindication 6 8 Daily maintenance 6 9 Trouble shooting 7 10 Storage and transportation 8 11 After service 8 12 Environmental protection 8 13 European authorized representative 8 14 Symbol instruction 8 15 EMC Declaration of conformity 9 16 Statement 12 17 Dental caries testing principle 13 ...

Page 3: ...ttery A full charge can be used for more than 400 times continuously under 10s working time mode 5 The check mode is used to detect dental calculus dental caries and cracked tooth 1 2 Principle and Application 1 2 1 X Cure adopts the principle of ray radiation to solidify the light sensitive resin by shooting at it in a short time 1 2 2 This product is used for dentistry It has the function of acc...

Page 4: ...2 ...

Page 5: ...in fuse L250V T3 15A 3 5 Light source 3 5 1 10W high power blue light LED 3 5 2 Wave length 385nm 515nm 3 5 3 Class class I 3 5 4 AEL 9 10 3J 3 5 5 check method When operate machine properly LED luminous means LED is in good condition 3 5 6 The wavelength of our curing light machine can match with dental resin material which are commonly used on clinical such as 3M and Dentsply 3 5 7 385nm to 515 ...

Page 6: ...r to the DC 15V input plug of the pedestal then put the main unit into the pedestal Please pull out the adapter after charging 5 Operation 5 1 Press the mode button to set the working mode the corresponding indicator will on when a mode set 5 2 High 2300 2500 mw cm2 5 3 Soft 1000 1200 mw cm2 5 4 Check 200 400mw cm2 5 5 Press the time button to set the solidifying time High 1S 2S 3S Soft 5S 10S 15S...

Page 7: ...ss power button 5 13 The effective light intensity of this equipment is much more higher than Halogen Lamp The solidified depth of the curing light composites resin for 10 seconds will not less than 4mm 6 Precaution 6 1 The optical fiber should be autoclaved under the high temperature of 134 and pressure 0 22 Mpa 6 2 Please recharge the battery at least 4 hours before first time usage 6 3 During o...

Page 8: ...terference the device should be installed at the medical site which meet the requirement of EMC WARNING The adapter should be connected to the socket which is easy for operator to touch WARNING over heat scorching the deivce cannot be used for 20s continuously WARNING High temperature burn The machine can not direct shining to the skin tissue like lips and mucosa WARNING Check mode Prohibited for ...

Page 9: ...op or us Light intensity is weak There is resin on the top of the optical fiber 1 Clean the resin 2 Change a new optical fiber T h e e q u i p m e n t d o e s n t c h a rg e when the adapter is connected 1 The adapter is not connected well 2 The charging point is impurity 3 Faulty of adapter or incompatible 1 Reconnect 2 Clean by the alcohol 3 Please contact our special repair shop or us Effective...

Page 10: ...ing transportation Handle with care 10 5 Don t put it together with dangerous articles during transportation 10 6 Keep it away form the sun rain or snow during transportation 11 After service From the date this equipment has been sold base on the warranty card we will repair this equipment free of charge if it has quality problems please refer to the warranty card for the warranty period 12 Enviro...

Page 11: ...ion electromagnetic emissions The models X Cure are intended for use in the electromagnetic environment specified below The customer or the user of the models X Cure should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The models X Cure use RF energy only for its internal function Therefore its RF emission...

Page 12: ...mmercial or hospital environment Surge lEC 61000 4 5 1 kV line to line 2 kV line to earth 1 kV line to line Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip i...

Page 13: ...cati on equipment Refer to table 9 of IEC 60601 1 2 2014 P o r t a b l e a n d m o b i l e R F communications equipment should be used no closer to any part of the models X Cure including cables than the r e c o m m e n d e d s e p a r a t i o n distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 3 5 V1 P1 2 d 1 2 P1 2 80 MHz to 80...

Page 14: ...interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the models X Cure are recommended below according to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150kHz to 80MHz d 1 2 P1 2 80MHz to 800MHz d 1 2 P1 2 800MHz to 2 ...