5
Figure 4 Charging
7) Gutta-percha loading slot:
Note: Only one gutta-percha can be loaded into it for a time.
8) Thermal Protector Cap:
The design of Thermal Protector Cap is to protect the oral soft tissue
and lip from scalding.
Note: Before being used to different patients or before each use, please
clean, disinfect and sterilize the Thermal Protector Cap.
9) Gutta Percha Injecting Needle
Before use, the Gutta Percha Injecting Needle and the Fi-G must be
connected and tightened with the wrench provided by our company to
prevent the needle from falling off or the gutta-percha leakage during use.
But do not to screw too tight. Here we provide a variety of needles with
different sizes (Please refer to Table 2), and the needles can be pre-bent
depending on the application. ( Note: Store unused Gutta Percha Injecting
Needles in a sealed environment, as the Gutta Percha Injecting Needle
are made of silver and may discolor due to oxidation caused by long-
term exposure to air ). Please use the wrench provided by the company to
connect, disassemble and pre-bend the Gutta Percha Injecting Needle.
Note: When replace the Gutta Percha Injecting Needle, please first
power off and wait for 5 minutes. Only after the Hot Melting and