13
the needle, the needle may break. While the gutta-percha is still hot, use a
medical vertical presser to squeeze down. If there are bubbles in the root
canal, use a small amount of material to fill the root canal for many times.
Use a little more material for each filling and use vertical presser to press
it down.
5
.
Replacement of gutta-percha
When the trigger is pushed forward to push the push ram to make a
"click" sound. it indicates that the gutta-percha in the Hot Melting and
Filling Instrument has been used up. And it is necessary to load a new
gutta-percha stick in time. When loading another gutta-percha stick,
make sure the Hot Melting and Filling Instrument has cooled to room
temperature. When the previous gutta-percha has been completely
squeezed out, then according to step 2, re-select the appropriate gutta-
percha for loading.
Warning:
Do not replace the gutta-percha stick in the heated state, otherwise it
may cause scalding or damage the Hot Melting and Filling Instrument.
6. After operation, the remaining materials in the heating chamber
must be cleaned, and the relevant accessories must be cleaned, disinfected
and sterilized. For details, see Chapter 8.
6 Charging instruction
6.1 Use corresponding charging base for charging: Connect the power
adapter to the charging base, and connect to power supply. And then
correctly place the Hot Melting and Filling Instrument in the charging
base. When the Hot Melting and Filling Instrument is not placed on the
charging base, the indicator will flashes in yellow and green alternately.
When the Hot Melting and Filling Instrument is placed on the charging
base, if the charging is being charged, the yellow indicator will be on
constantly. When the battery is full, the yellow indicator will be off and
the green indicator will be on constantly. Under normal situation, the
charging takes about 2.5h.
6.2 The battery used in this product has no memory and can be used at
any time or charged at any time.
6.3 Before first use of this device, please charge it at least for 3 hours.
Warning:
Only unplug the adapter to disconnect from the network power.