21
Percha material extrusion, and will Pushing Ram, Thermal Protector
Cap, Gutta Percha Injecting Needle Protector and Gutta Percha Injecting
Needle after unloaded, please use the Cleaning Brush provided from the
back of the host in them, and come out from the front of the host, clean
at least 3 times, until the host fever intracavitary residual Gutta Percha
material is clean.
• After each use, wipe the surface of the main engine and charging base
as well as the cleaning brush and Pushing Ram with a clean soft cloth
soaked in purified water (distilled or deionized) or a clean disposable wet
paper towel, and repeat for at least 3 times.
8.12 Daily maintenance
When the device is not used, please turn off the power and unplug the
power supply plug.
If the Hot Melting and Filling Instrument is in a low battery state for a
long time, the service life of battery will be shortened. Please charge it in
time if the battery level is low. When the device is not used, please charge
it for 1 hour once a month.
Warning:
Equipment and all accessories shall not be maintained during use
8.13 Repair of device
This product does not contain self-repairing spare parts. If there
is any abnormality in the equipment, please contact our company
for maintenance and do not disassemble without authorization. With
our company's consent, we will provide circuit diagrams, component
part lists, descriptions, calibration instructions to assist to SERVICE
PERSONNEL in parts repair.