5
commissioned in strict conformity with the EMC information provided
in this instruction manual. Specifically, do not use the device close to
fluorescent lamp, radio transmitters and remote controls.
1.10.5 Long time use of the device will lead to overheating of the
micro motor, let it cold down that use. If the motor handpiece overheats
too persists, contact your distributor.
1.10.6 The USB port of the base must only be connected to USB port
of the handpiece through the USB wire. Never use it for other purposes.
1.10.7 Over-heat scorching: the handpiece cannot be used for 10
minutes continuously.
Contra-angle
1.10.8 Only use the original contra-angle. Do not use any other
contra-angle or other reduction rate other than original one.
1.10.9 Never press the contra-angle push button when the motor
handpiece is running or if it is coming to a stop. This will lead to
detachment of the instrument or cause the pushbutton to overheat.
1.10.10 Never remove the contra-angle from the motor handpiece
during operation.
1.10.11 Only use undamaged root canal instruments. Please refer to
the information provided by the manufacturer.
1.10.12 Only insert the instrument when the contra-angle is
stationary.
1.10.13 Never place your fingers on the moving parts of the
instrument while it is running or coming to a stop.
1.10.14 Before treatment, check the contra-angle for any damage or
loose parts.
Root canal instruments
1.10.15 Before use, make sure the instrument is securely locked in
place.
1.10.16 Never use continuous rotary instruments in Safety Glide Path
mode.
1.10.17 Never use Safety Glide Path instruments in rotary mode.
1.10.18 Use the torque and speed settings recommended by the
instrument manufacturer.
1.10.19 Replacement of lithium batteries by inadequately trained
personnel could result in a HAZARD,so please contact local distributors
to replace the battery if necessary.
1.10.20 The adapter plug can be used to disconnect from the network