34
2. Check the product. If it is obviously damaged, smashed, detached,
corroded or bent, it must be scrapped and not allowed to continue to be
used.
3. Check the product. If the accessories are found to be damaged,
please replace it before use. And the new accessories for replacement
must be cleaned, disinfected and dried.
4. If the service time (number of times) of the product reaches the
specified service life (number of times), please replace it in time.
7.4.6.2Maintenance
Oil lubrication of sterilized and dried products.
The nozzle of cleaning lubricant is aligned with the air intake hole at
the end of thecontra angle to inject oil for 1-2 seconds.
7.4.7 Packaging
Install the disinfected and dried product and quickly package it in a
medical sterilization bag (or special holder, sterile box).
Notes:
a) The package used conforms to ISO 11607;
b) It can withstand high temperature of 138 °C and has sufficient
steam permeability;
c) The packaging environment and related tools must be cleaned
regularly to ensure cleanliness and prevent the introduction of
contaminants;
d) Avoid contact with parts of different metals when packaging.
7.4.8 Sterilization
Use only the following steam sterilization procedures (fractional pre-
vacuum procedure*) for sterilization, and other sterilization procedures
are prohibited:
·The steam sterilizer complies with EN13060 or is certified according
to EN 285 to comply with EN ISO 17665;