14
10 Display the current remaining battery power of the base
Fully charged
The remaining battery power is about 70%
The remaining battery power is about 30%
The battery is very low, please charge in time
11 Display the current Safety Glide Path mode file system number
12 Reverse angle, adjustable from 20° to 400°
13 Forward angle, adjustable from 20° to 400°
14 Apex bar Display apex bar
2.10 ATR Mode
ATR: Adaptive Torque Reverse function.
Normal continuous forward rotation, the forward angle can be
stepped by 10°, the angle is set between 20°-400°, and the reverse angle
defaults to 90°. When the load of the file is greater than the set torque
limit, the file will start to rotate alternately at the set angle.
Trigger torque: 0.4Ncm-4Ncm
Speed: 100rpm-500rpm
2.11 Standby mode
If the device is not used for 3 minutes, the motor handpiece will
automatically shut down. In shutdown mode, press the handle button,
you can boot instantly.