Guilin Woodpecker Medical Instrument Endo Radar Pro Instruction Manual Download Page 27

25

1) Approx 2mm to apical foramen

2) Approx 1mm to apical foramen

3) Approx 0.5mm to apical foramen

4) Apex (apical foramen)

  WARNNING

In the combined length determination mode,If the running time 

reaches 1 minute, please allow the device to cool down for 1 minute and 

start again the motor handpiece.

6 Trouble shooting 

When trouble is found, check the following points before contacting 

your distributor. If none of these are applicable or the trouble is not 

remedied even after action has been taken, the product may have failed. 

Contact your distributor.

Summary of Contents for Endo Radar Pro

Page 1: ...Endo Radar Pro Endo Motor INSTRUCTION MANUAL...

Page 2: ...ined length determination mode 23 6 Trouble shooting 25 7 Cleaning Disinfection and Sterilization 27 8 Storage maintenance and transportation 36 9 Environmental protection 36 10 After service 36 11 Eu...

Page 3: ...t is a cordless endo motor with root canal measurement capability It can be used as a endo motor for preparation and enlargement of root canals or device for measuring canal length It can be used to e...

Page 4: ...1 Structure Endo Radar Pro is composed of Motor handpiece Contra angle measuring wire USB wire Power adapter etc 1 3 2 Product accessories Motor handpiece Contra angle Power adapter Base Measuring wir...

Page 5: ...aration and enlargement of root canals or device for measuring canal length 1 5 2 The device must be operated in hospital and clinic by the qualified dentists 1 6 Contraindications In cases where a pa...

Page 6: ...m battery in base 11 1V 2600mAh Lithium battery in motor handpiece 3 7V 1200mAh 1 8 2 Power adapter Input 100V 240V 50 60Hz 800mA Output DC15V 1 6A 1 8 3 Torque rang 0 4Ncm 5 0Ncm 4mNm 50mNm 1 8 4 Spe...

Page 7: ...ment of the instrument or cause the pushbutton to overheat 1 10 10 Never remove the contra angle from the motor handpiece during operation 1 10 11 Only use undamaged root canal instruments Please refe...

Page 8: ...ve the accidental crash of the base or accidental failure of the touch button 2 2 Installing and removing the contra angle 2 2 1 Installing Align any locating pin of the contra angle with the position...

Page 9: ...e power off beforehand 2 3 Installing and removing the protective silicon cover 2 3 1 Installing Put the protective silicon cover onto the contra angle 2 3 2 Removing When removing the protective sili...

Page 10: ...ve until there is no obvious wrinkle After installing the disposable isolation sleeve wrap the barrier film around the handpiece surface After that clean and disinfect the surface of the handpiece Ref...

Page 11: ...h Mode decrease the angle 9 SPEED Increase the motor rotation speed 10 SPEED Decrease the motor rotation speed 11 TORQUE Increase the torque 12 TORQUE Decrease the torque 13 FILE Positive selection ke...

Page 12: ...0mm to apex 0 0mm 4 5 SPEED Select 6 En Ap Select or or enable or disable or enable or disable or or Enable auto start The file will start rotating when it is inserted the canal Enable auto start and...

Page 13: ...on the display must always match the file in use 2 8 3 Torque and speed values are subject to change by the file manufacturers without notice Therefore the preset values in the library must be checked...

Page 14: ...12 1 SYSTEM Display the selected file system 2 FILE Display the selected file type 3 SPEED Display the rotation speed applied only to continuous rotation mode...

Page 15: ...g Counterclockwise continuously rotating Safety Glide Path motion mode 8 Wireless communication connection status indication The motor handpiece is connected to the base wirelessly Wireless communicat...

Page 16: ...ex bar Display apex bar 2 10 ATR Mode ATR Adaptive Torque Reverse function Normal continuous forward rotation the forward angle can be stepped by 10 the angle is set between 20 400 and the reverse ang...

Page 17: ...otor handpiece in power on state 3 2 2 After the base is powered on in the power on state of handpiece if the base wireless communication connection icon is lit the motor handpiece had established wir...

Page 18: ...n 1 in the picture The current load is 50 of the preset load b when it shows location 2 in the picture The current load is 80 of the preset load c when it shows location 2 in the picture The current l...

Page 19: ...torque by the contra angle Calibration is recommended when using a new other contra angle or after an extended period of operation as the running properties can change with usage cleaning and sterili...

Page 20: ...ess induction charging of the motor handpiece After the motor handpiece is fully charged the wireless charging indicator would be off WARNNING a Do not open the device or change the battery That may c...

Page 21: ...ice away 4 1 2 Insert the file clip and lip hook respectively into the two sockets of the measuring wire Picture 1 Attention Be sure not to pull the wire when inserting or pulling out the measuring wi...

Page 22: ...damaged should be replaced Picture 3 Picture 4 a Picture 4 b Picture 4 c 4 3 Explanations on the interfaces displayed a The screen displays the front region of the apical foramen by instruction bars P...

Page 23: ...5 Operation instruction 4 5 1 Please let the measure wire insert the base Then staring up the power button Next press the En Ap button And you can enter into the apex mode 4 5 2 When the device is st...

Page 24: ...t will make the result of measurement inaccurate d Please reference the Picture 8 a to grip the needle file If as Picture 8 b it can t 4 5 5 When the file reaches the apex adjust the rubber piece set...

Page 25: ...eatment WARNNING Measuring wire cannot be clean by high temperature and high pressure 5 Combined length determination mode For a combined length determination 5 1 Base installing 5 1 1 Insert the meas...

Page 26: ...Ap button to select the Combined length determination mode 5 3 2 Select file system 5 3 3 Combined length determination mode parameter setting See 2 7 5 4 Connection testing Test every time before usi...

Page 27: ...the running time reaches 1 minute please allow the device to cool down for 1 minute and start again the motor handpiece 6 Trouble shooting When trouble is found check the following points before cont...

Page 28: ...n has been interrupted calibrate the motor handpiece again to rule out the possibility of a motor fault 2 Clean and lubricate the contra angle 3 Start the calibration procedure again Motor handpiece i...

Page 29: ...otective equipment gloves glasses mask 7 2 5 The user is responsible for the sterility of the product for the first cycle and each further usage as well as for the usage of damaged or dirty instrument...

Page 30: ...7 3 2 1 Manual disinfection steps 1 Soak the dry soft cloth with 75 alcohol 2 Wipe all surfaces of headpiece charger base and other components with a wet soft cloth for at least 3 minutes 3 Wipe the...

Page 31: ...re and time specified by the disinfectant manufacturer 7 4 The cleaning disinfection and sterilization of contra angle lip hook file clip protective silicon cover touch probeare as follow Unless other...

Page 32: ...nse away the dirt on the surface of handpiece with pure water or distilled water deionized water 2 Dry the products with a clean soft cloth and place it in a clean tray Notes a The water used here mus...

Page 33: ...mated cleaning The cleaner is proved to be valid by CE certification in accordance with EN ISO 15883 There should be a flushing connector connected to the inner cavity of the product The cleaning proc...

Page 34: ...are not allowed to contact each other 2 Use a suitable rinsing adaptor and connect the internal water lines to the rinsing connection of the washer disinfector 3 Start the program 4 After the program...

Page 35: ...e disinfector 7 4 5 Drying If your cleaning and disinfection process does not have an automatic drying function dry it after cleaning and disinfection Methods 1 Spread a clean white paper white cloth...

Page 36: ...contra angle to inject oil for 1 2 seconds 7 4 7 Packaging Install the disinfected and dried product and quickly package it in a medical sterilization bag or special holder sterile box Notes a The pac...

Page 37: ...ture plasma sterilization The manufacturer assumes no responsibility for the procedures that have not been recommended If you use the sterilization procedures that have not been recommended please adh...

Page 38: ...8 2 Maintenance 8 2 1 This device do not include accessories for repair usage the repair should be carried out by authorized person or authorized after service center 8 2 2 Keep the equipment in a dry...

Page 39: ...low Instructions for Use CE marked product Appliance compliance WEEE directive Authorised Representative in the EUROPEAN COMMUNITY 13 Statement All rights of modifying the product are reserved to the...

Page 40: ...ns CISPR 11 Group 1 The model Endo Radar Pro uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equ...

Page 41: ...s and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 5 UT 95 dip in UT for 1 cycle 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 250 cycles 5 UT...

Page 42: ...n the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P1 2 d 2 P1 2 d 1 2 P1 2 80 MHz to 800 MHz d 2 3...

Page 43: ...ess than 3V m Table 4 Recommended separation distances between portable and mobile RF communications equipment and the model Endo Radar Pro Recommended separation distances between portable and mobile...

Page 44: ...the transmitter where P is the maximum output power rating of the transmitter in watts W accordable to the transmitter manufacturer NOTE I At 80 MHz and 800 MHz the separation distance for the higher...

Page 45: ...ZMN SM 408 V1 3 20210524...

Reviews: