GUIDO RAYOS X, S.A.
NESTORET 5050
30
MAN-011
Feb. 01
Ed. 2 / Rev. 1
The NESTORET 5050 Incubator has the capability to read signals
generated by an Air Temperature sensor, by a Skin Temperature sensor,
by a Humidity sensor, by an Oxygen Concentration sensor, by a Heating
Device Temperature sensor and the Fan Operation by HALL effect.
The Air Temperature Sensor is amplified by
U7
being possible to adjust
the
P5
gain and the
P13
offset.
The Skin Temperature sensor is isolated from the circuit by means the
U8
isolation amplifier that provides a 2500 Vpp isolation. At the output of
this amplifier the
U9
circuit amplifies the sensor isolated signal being
possible to adjust the gain and offset by the
P14
and
P8
respectively.
The Humidity sensor is a capacitive sensor with signal conditioning. It
makes a correction in the relative humidity depending on the temperature.
The humidity correction factor can be adjust by the
P1
potentiometer of
the humidity board. Gain level is adjustable by
P2
.
The signal generated by the Oxygen Concentration sensor is amplified by
U16
, being possible to adjust the gain and offset by
P12
and
P7
respectively.
The Heating Device Temperature sensor is amplified by
U13
being
possible to adjust the
P10
gain and
P11
offset.
When fan blades rotation, the HALL effect sensor generates a pulse.
Those pulses are integrated by
C19
,
D8
,
R66
and
C21
, having at
TP9
a
proportional power to the fan rotating speed.
The
J11
and
J12
Connectors are free to receive any other analogic signal
that could be used on future implements. At present these two inputs are
deactivated in the activation/deactivation panel from the service program.
5. Power Drivers
The heating device and the fan are activated by their corresponding
driver. The signal coming from the Microcontroller is isolated by the
OPT1-OPT2
Optocoupler that feeds the
TRC1-TRC2
Triac supplying
power to the heater and to the fan. The heater is activated and deactivated
upon requirement of the system. The fan is activated/deactivated when
pressing
ON/OFF
key.
IV.2. ADJUSTMENTS
Here below are described the adjustment instructions to perform on the NESTORET
5050 in case of replacement of some of the components, such as sensors, circuits, etc.,
due to a failure.
To perform the adjustments properly the following instruments are required:
•
Digital Voltmeter (DVM).
•
Temperature and Oxygen Concentration Simulator (SIM50TO).
•
Humidity Concentration Simulator (SIM50HR).
Summary of Contents for NESTORET 5050
Page 1: ...GUIDO RAYOS X S A NESTORET 5050 1 MAN 011 Feb 01 Ed 2 Rev 1...
Page 4: ...GUIDO RAYOS X S A NESTORET 5050 4 MAN 011 Feb 01 Ed 2 Rev 1...
Page 6: ...GUIDO RAYOS X S A NESTORET 5050 6 MAN 011 Feb 01 Ed 2 Rev 1...
Page 12: ...GUIDO RAYOS X S A NESTORET 5050 12 MAN 011 Feb 01 Ed 2 Rev 1...
Page 42: ...GUIDO RAYOS X S A NESTORET 5050 42 MAN 011 Feb 01 Ed 2 Rev 1...
Page 44: ...GUIDO RAYOS X S A NESTORET 5050 44 MAN 011 Feb 01 Ed 2 Rev 1...
Page 47: ...GUIDO RAYOS X S A NESTORET 5050 47 MAN 011 Feb 01 Ed 2 Rev 1...
Page 49: ...GUIDO RAYOS X S A NESTORET 5050 49 MAN 011 Feb 01 Ed 2 Rev 1...
Page 53: ...GUIDO RAYOS X S A NESTORET 5050 53 MAN 011 Feb 01 Ed 2 Rev 1...
Page 59: ...GUIDO RAYOS X S A NESTORET 5050 59 MAN 011 Feb 01 Ed 2 Rev 1...
Page 60: ...GUIDO RAYOS X S A NESTORET 5050 60 MAN 011 Feb 01 Ed 2 Rev 1...