GUIDO RAYOS X, S.A.
NESTORET 5050
19
MAN-011
Feb. 01
Ed. 2 / Rev. 1
OX CALIBR (21 %)
PRESS ENTER
By pressing
ENTER
, the Oxygen Sensor is already calibrated with a new
reference.
New screen displayed ;
AIR=__._ ºC
OX: 1=PRG, 2=CAL
To set the
Relative Humidity
procedure is the same we have explained to
Oxygen values, but selecting 2=HUMID, instead of 1=OXYG.
II.5. ALARMS
The NESTORET 5050 Alarm System informs to the medical attendant of any incidence
may occur. The alarm is visible and audible.
NOMENCLATURE OF SYMBOLS IN ALARM
MESSAGES
+ Excess temperature alarm
-
Failing temperature alarm
= Temperautre O.K.
*
Desactive sensor
A) AIR CONTROL MODE
1. SENSOR ALARM:
In case of air sensor failure or it gets disconnected or out of range; the
SENSOR Alarm will activate the corresponding LED and buzzer, the
alarm cannot be silenced. The LCD will inform to the attendant
displaying ,
* ALARM *
AIR SENSOR
The Incubator is now out of service. Call to any authorized Service
engineer.
Summary of Contents for NESTORET 5050
Page 1: ...GUIDO RAYOS X S A NESTORET 5050 1 MAN 011 Feb 01 Ed 2 Rev 1...
Page 4: ...GUIDO RAYOS X S A NESTORET 5050 4 MAN 011 Feb 01 Ed 2 Rev 1...
Page 6: ...GUIDO RAYOS X S A NESTORET 5050 6 MAN 011 Feb 01 Ed 2 Rev 1...
Page 12: ...GUIDO RAYOS X S A NESTORET 5050 12 MAN 011 Feb 01 Ed 2 Rev 1...
Page 42: ...GUIDO RAYOS X S A NESTORET 5050 42 MAN 011 Feb 01 Ed 2 Rev 1...
Page 44: ...GUIDO RAYOS X S A NESTORET 5050 44 MAN 011 Feb 01 Ed 2 Rev 1...
Page 47: ...GUIDO RAYOS X S A NESTORET 5050 47 MAN 011 Feb 01 Ed 2 Rev 1...
Page 49: ...GUIDO RAYOS X S A NESTORET 5050 49 MAN 011 Feb 01 Ed 2 Rev 1...
Page 53: ...GUIDO RAYOS X S A NESTORET 5050 53 MAN 011 Feb 01 Ed 2 Rev 1...
Page 59: ...GUIDO RAYOS X S A NESTORET 5050 59 MAN 011 Feb 01 Ed 2 Rev 1...
Page 60: ...GUIDO RAYOS X S A NESTORET 5050 60 MAN 011 Feb 01 Ed 2 Rev 1...