Hexylon High-performance portable meter
10
Power supply
A DC adapter is provided to power and charge the
meter. Plug the adapter into a properly grounded
electrical supply and the power connector on the
side of the unit.
When external power is supplied, the battery
management system automatically controls the
charging process.
A battery icon indicates the charge status of the
battery.
When the battery is fully charged, the battery icon
is completely filled. As the battery discharges, the
amount the icon is filled decreases in steps.
From a fully discharged state, a full charge takes
approximately between 3 and 4 hours if the
HEXYLON is turned off. When the meter is turned
on, the charge takes more time.
When the meter is off while the battery is charging,
the ON/OFF button red led blinks. When the
battery is totally charged the red light remains red.
A 1 hour charge will provide battery for around two
working hours.
The charge management system will detect
various conditions preventing charging, such as a
battery that is over a safe temperature.
About the battery
Important:
If the equipment is going to be stored for a while,
it is recommended to take the battery off and store
them both separately.
To maximize battery life:
Allow fully discharging the battery.
The battery can be charged correctly attached
to the device and using the supplied DC adapter.
Or separately using the the DC adapter supplied
with the Hexylon.
For long term storage, take the battery pack
off, and keep the device and the battery pack
separately at room temperature, or about 25º C.
Start with a charged battery and re-charge the
battery every 2 to 3 months.
Battery replacement:
It is recommended to use only the battery packs
supplied by the manufacturer to replace the
battery, following these instructions:
Disconnect the meter from the power supply:
Turn the Hexylon off: